Revisi<strong>on</strong>: <str<strong>on</strong>g>Final</str<strong>on</strong>g> <str<strong>on</strong>g>Report</str<strong>on</strong>g><br />

Date: May 25, 2012<br />

indentati<strong>on</strong> in the comparis<strong>on</strong>. There are 9 areas of measurement <strong>on</strong> each block. The<br />

rest of the segments <strong>on</strong> the block surface are reserved as alternative places for trial and<br />

error measurements<br />

Figure 1 Indentati<strong>on</strong> positi<strong>on</strong><br />

The plastic sheet indicating the indentati<strong>on</strong> positi<strong>on</strong> of each participant will be<br />

enclosed with the artifact sets. It shall be placed <strong>on</strong> the artifact block in order to ensure<br />

the indentati<strong>on</strong> positi<strong>on</strong> before making the indentati<strong>on</strong>.<br />

5. Measurement of Artifacts<br />

The primary Rockwell hardness machines used for the comparis<strong>on</strong> must be passed<br />

the verificati<strong>on</strong> according to ISO 6508-3 [1]. Before making the indentati<strong>on</strong> <strong>on</strong> each<br />

block, the testing cycle must be adjusted to comply with ISO 6508-3. Then start to<br />

measure the hardness value of artifact blocks by making 9 indentati<strong>on</strong>s each block <strong>on</strong><br />

the designated area <strong>on</strong> the indentati<strong>on</strong>-positi<strong>on</strong>ing sheet.<br />

Each participant must give a characteristics descripti<strong>on</strong> of your primary Rockwell<br />

hardness machine, your primary indenter, testing c<strong>on</strong>diti<strong>on</strong> and envir<strong>on</strong>mental<br />

informati<strong>on</strong>. Also laboratory must report the measurement results for each indentati<strong>on</strong><br />

and the uncertainty of measurement of Rockwell hardness scale A and Rockwell<br />

hardness scale B, and uncertainty budget of measurement.<br />