Revisi<strong>on</strong>: <str<strong>on</strong>g>Final</str<strong>on</strong>g> <str<strong>on</strong>g>Report</str<strong>on</strong>g><br />

Date: May 25, 2012<br />

Data report forms<br />

<str<strong>on</strong>g>Report</str<strong>on</strong>g> Form 1 (For Rockwell hardness scale B)<br />

Metrology informati<strong>on</strong><br />

Model of hardness primary machine 8150TK (Note 1)<br />

Serial number of hardness primary machine 093925 (Note 1)<br />

Manufacturer of hardness primary machine Indentec (Note 1)<br />

<strong>Hardness</strong> indicati<strong>on</strong> system Length transducer<br />

Test force system Deadweight<br />

Relative error of preliminary test force 0.0 %<br />

Relative uncertainty (2s) of preliminary test force 1.0 %<br />

Relative error of total test force 0.0 %<br />

Relative uncertainty (2s) of total test force 0.1 %<br />

Uncertainty of depth measuring device 3 um<br />

The calibrati<strong>on</strong> and measurement capability (CMC) 1.5 HRBS (Note 2)<br />

Note 1: The machine is not a primary hardness machine, but a commercial hardness machine.<br />

Note 2: This CMC has not been submitted to <str<strong>on</strong>g>APMP</str<strong>on</strong>g> TCM for approval yet.<br />

Velocity of the indenter when reaching the surface NA*<br />

Holding time of preliminary test force 6.4 s<br />

Test force loading speed 88 um/s<br />

Total test force durati<strong>on</strong> time 4.7 s<br />

Time before reading permanent indentati<strong>on</strong> depth NA*<br />

Measurement place SCL H<strong>on</strong>g K<strong>on</strong>g<br />

Measurement date 3 August 2010<br />

Deformati<strong>on</strong> of frame (um)<br />

Testing c<strong>on</strong>diti<strong>on</strong><br />

* SCL hardness machine equips with a n<strong>on</strong>-motorised leadscrew loading system with upward moti<strong>on</strong> driven manually.<br />

Envir<strong>on</strong>ment<br />

Envir<strong>on</strong>ment of comparis<strong>on</strong> measurement<br />

(20 ± 1) o C<br />

(50 ± 8) %RH<br />

(991 ± 3) hPa<br />

Deformati<strong>on</strong> of frame verificati<strong>on</strong><br />

1 2 3<br />

0.7 0.9 0.8<br />