Revisi<strong>on</strong>: <str<strong>on</strong>g>Final</str<strong>on</strong>g> <str<strong>on</strong>g>Report</str<strong>on</strong>g><br />
Date: May 25, 2012<br />
Data report forms<br />
<str<strong>on</strong>g>Report</str<strong>on</strong>g> Form 1 (For Rockwell hardness scale B)<br />
Metrology informati<strong>on</strong><br />
Model of hardness primary machine 8150TK (Note 1)<br />
Serial number of hardness primary machine 093925 (Note 1)<br />
Manufacturer of hardness primary machine Indentec (Note 1)<br />
<strong>Hardness</strong> indicati<strong>on</strong> system Length transducer<br />
Test force system Deadweight<br />
Relative error of preliminary test force 0.0 %<br />
Relative uncertainty (2s) of preliminary test force 1.0 %<br />
Relative error of total test force 0.0 %<br />
Relative uncertainty (2s) of total test force 0.1 %<br />
Uncertainty of depth measuring device 3 um<br />
The calibrati<strong>on</strong> and measurement capability (CMC) 1.5 HRBS (Note 2)<br />
Note 1: The machine is not a primary hardness machine, but a commercial hardness machine.<br />
Note 2: This CMC has not been submitted to <str<strong>on</strong>g>APMP</str<strong>on</strong>g> TCM for approval yet.<br />
Velocity of the indenter when reaching the surface NA*<br />
Holding time of preliminary test force 6.4 s<br />
Test force loading speed 88 um/s<br />
Total test force durati<strong>on</strong> time 4.7 s<br />
Time before reading permanent indentati<strong>on</strong> depth NA*<br />
Measurement place SCL H<strong>on</strong>g K<strong>on</strong>g<br />
Measurement date 3 August 2010<br />
Deformati<strong>on</strong> of frame (um)<br />
Testing c<strong>on</strong>diti<strong>on</strong><br />
* SCL hardness machine equips with a n<strong>on</strong>-motorised leadscrew loading system with upward moti<strong>on</strong> driven manually.<br />
Envir<strong>on</strong>ment<br />
Envir<strong>on</strong>ment of comparis<strong>on</strong> measurement<br />
(20 ± 1) o C<br />
(50 ± 8) %RH<br />
(991 ± 3) hPa<br />
Deformati<strong>on</strong> of frame verificati<strong>on</strong><br />
1 2 3<br />
0.7 0.9 0.8<br />
Page 70 of 77