Revisi<strong>on</strong>: <str<strong>on</strong>g>Final</str<strong>on</strong>g> <str<strong>on</strong>g>Report</str<strong>on</strong>g><br />

Date: May 25, 2012<br />

Uncertainty budget for<br />

60 HRA<br />

No Quantity<br />

Estimated<br />

valus<br />

Unit<br />

Stadard<br />

uncertatinty<br />

Sensitivity<br />

coefficient<br />

Uncertainty<br />

c<strong>on</strong>tributi<strong>on</strong><br />

X i x i<br />

xi<br />

ux<br />

<br />

C i u i (H )<br />

1 Preliminary test force 0.09 N 0.05 0.1342 0.0071<br />

2 Total test force 0.16 N 0.09 -0.0574 -0.0053<br />

3 Permanent depth of indentati<strong>on</strong> 0.2 ㎛ 0.12 -0.5 -0.0577<br />

4 Preliminary test force dwell time 0.5 s 0.29 0.0214 0.0062<br />

5 Indentati<strong>on</strong> velocity 6 ㎛/s 3.46 -0.0014 -0.0048<br />

6 Total test force dwell time 0.5 s 0.29 -0.066 -0.0191<br />

7 indenter angle 0.04 ° 0.02 0.18 0.0042<br />

8 Indenter radius 3.1 ㎛ 1.79 0.058 0.1038<br />

9 Deformati<strong>on</strong> of frame 0.2 ㎛ 0.12 0.5 0.0577<br />

10<br />

L<strong>on</strong>g-term stability of primary<br />

hardness testing machine<br />

0.23 HRA 0.13 1<br />

0.1328<br />

11 Combined uncertainty(HRA) 0.189<br />

12 Expanded uncertainty(HRA), k=2 0.38<br />

Uncertainty budget for<br />

70 HRA<br />

No Quantity<br />

Estimated<br />

valus<br />

Unit<br />

Stadard<br />

uncertatinty<br />

Sensitivity<br />

coefficient<br />

Uncertainty<br />

c<strong>on</strong>tributi<strong>on</strong><br />

X i x i<br />

xi<br />

ux<br />

<br />

C i u i (H )<br />

1 Preliminary test force 0.09 N 0.05 0.097 0.0052<br />

2 Total test force 0.16 N 0.09 -0.0437 -0.0040<br />

3 Permanent depth of indentati<strong>on</strong> 0.2 ㎛ 0.12 -0.5 -0.0577<br />

4 Preliminary test force dwell time 0.5 s 0.29 0.0104 0.0030<br />

5 Indentati<strong>on</strong> velocity 6 ㎛/s 3.46 0.0021 0.0073<br />

6 Total test force dwell time 0.5 s 0.29 -0.05 -0.0144<br />

7 indenter angle 0.04 ° 0.02 0.18 0.0042<br />

8 Indenter radius 3.1 ㎛ 1.79 0.058 0.1038<br />

9 Deformati<strong>on</strong> of frame 0.2 ㎛ 0.12 0.5 0.0577<br />

10<br />

L<strong>on</strong>g-term stability of primary<br />

hardness testing machine<br />

0.23 HRA 0.13 1<br />

0.1328<br />

11 Combined uncertainty(HRA) 0.188<br />

12 Expanded uncertainty(HRA), k=2 0.38<br />

Uncertainty budget for<br />

80 HRA<br />

No Quantity<br />

Estimated<br />

valus<br />

Unit<br />

Stadard<br />

uncertatinty<br />

Sensitivity<br />

coefficient<br />

Uncertainty<br />

c<strong>on</strong>tributi<strong>on</strong><br />

X i x i<br />

xi<br />

ux<br />

<br />

C i u i (H )<br />

1 Preliminary test force 0.09 N 0.05 0.0598 0.0032<br />

2 Total test force 0.16 N 0.09 -0.03 -0.0028<br />

3 Permanent depth of indentati<strong>on</strong> 0.2 ㎛ 0.12 -0.5 -0.0577<br />

4 Preliminary test force dwell time 0.5 s 0.29 -0.0006 -0.0002<br />

5 Indentati<strong>on</strong> velocity 6 ㎛/s 3.46 0.0056 0.0194<br />

6 Total test force dwell time 0.5 s 0.29 -0.034 -0.0098<br />

7 indenter angle 0.04 ° 0.02 0.18 0.0042<br />

8 Indenter radius 3.1 ㎛ 1.79 0.058 0.1038<br />

9 Deformati<strong>on</strong> of frame 0.2 ㎛ 0.12 0.5 0.0577<br />

10<br />

L<strong>on</strong>g-term stability of primary<br />

hardness testing machine<br />

0.20 HRA 0.12 1<br />

0.1155<br />

11 Combined uncertainty(HRA) 0.177<br />

12 Expanded uncertainty(HRA), k=2 0.35<br />

i<br />

i<br />

i<br />

3<br />

3<br />

3<br />