APMP.M.H-S3 Final Report Comparison on Hardness ... - BIPM
APMP.M.H-S3 Final Report Comparison on Hardness ... - BIPM
APMP.M.H-S3 Final Report Comparison on Hardness ... - BIPM
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
Revisi<strong>on</strong>: <str<strong>on</strong>g>Final</str<strong>on</strong>g> <str<strong>on</strong>g>Report</str<strong>on</strong>g><br />
Date: May 25, 2012<br />
Data report forms Page 2/5<br />
<str<strong>on</strong>g>Report</str<strong>on</strong>g> Form 1 (For Rockwell hardness scale B)<br />
Metrology informati<strong>on</strong><br />
Model of hardness primary machine SHT-31<br />
Serial number of hardness primary machine 310037<br />
Manufacturer of hardness primary machine Akashi<br />
<strong>Hardness</strong> indicati<strong>on</strong> system Laser Hologauge<br />
Test force system Lever<br />
Relative error of preliminary test force 0.166%<br />
Relative uncertainty (2) of preliminary test force 0.126%<br />
Relative error of total test force 0.025%<br />
Relative uncertainty (2) of total test force 0.030%<br />
Uncertainty of depth measuring device 0.12 ㎛<br />
The calibrati<strong>on</strong> and measurement capability (CMC) 0.6 HRBW<br />
Testing c<strong>on</strong>diti<strong>on</strong><br />
Velocity of the indenter when reaching the surface 25㎛/s ~ 37㎛/s<br />
Holding time of preliminary test force 3.0 sec<br />
Test force loading speed 5.0 sec<br />
Total test force durati<strong>on</strong> time 4.0 sec<br />
Time before reading permanent indentati<strong>on</strong> depth 1.0 sec<br />
Envir<strong>on</strong>ment of comparis<strong>on</strong> measurement<br />
Measurement place KRISS<br />
Measurement date Sep. 30, 2010<br />
Envir<strong>on</strong>ment 23 °C, 43% R.H.<br />
Deformati<strong>on</strong> of frame verificati<strong>on</strong><br />
Deformati<strong>on</strong> of frame (m)<br />
1 2 3<br />
-0.2 -0.2 -0.2<br />
Page 52 of 77