19.06.2013 Views

APMP.M.H-S3 Final Report Comparison on Hardness ... - BIPM

APMP.M.H-S3 Final Report Comparison on Hardness ... - BIPM

APMP.M.H-S3 Final Report Comparison on Hardness ... - BIPM

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

Revisi<strong>on</strong>: <str<strong>on</strong>g>Final</str<strong>on</strong>g> <str<strong>on</strong>g>Report</str<strong>on</strong>g><br />

Date: May 25, 2012<br />

Uncertainty budget of <strong>Hardness</strong> machine at 60 HRB 60<br />

<strong>Hardness</strong> value, HRB 59.6 3549.1 59.6 1<br />

Quantity Unit Estimated Standard Sensitivity Uncertainty<br />

error value uncertainty coefficient c<strong>on</strong>tributi<strong>on</strong><br />

Xi xi u(xi) ci ui(H) Preliminary test force, Fo N 0.020 0.011 0.102 0.001<br />

Total test force, F N 0.981 0.566 -0.084 -0.048<br />

Permanent depth of indentati<strong>on</strong>, h m 0.2 0.115 -0.500 -0.058<br />

Preliminary test force dwell time, to s 0.5 0.289 0.118 0.034<br />

Indentati<strong>on</strong> velocity, v mm/s -1<br />

0.3 0.173 0.002 0.000<br />

Total test force dwell time, t s 1 0.289 -0.017 -0.005<br />

Indenter ball diameter, d m 2 1.155 -0.007 -0.008<br />

Deformati<strong>on</strong> of frame, DF m 0.25 0.144 0.500 0.072<br />

Combined uncertainty, u c HRB Combined uncertainty, uc 0.110<br />

Expanded uncertainty, U e HRB Expanded uncertainty, Ue 0.220<br />

Uncertainty budget of hardness measurement at 60 HRB 60<br />

Quantity Unit Estimated Standard Sensitivity Uncertainty<br />

value uncertainty coefficient c<strong>on</strong>tributi<strong>on</strong><br />

X i x i u(x i) c i u i(H)<br />

<strong>Hardness</strong> machine, H phm HRB 59.6 0.110 1 0.110<br />

Readability,H res HRB 0 0.014 1 0.014<br />

Repeatability of measurement, H rep HRB 0 0.110 1 0.110<br />

Combined uncertainty, uc 0.156<br />

Measured hardness,H mh HRB 59.6 Expanded uncertainty, Ue 0.313<br />

Uncertainty budget of <strong>Hardness</strong> machine at 70 HRB 70<br />

<strong>Hardness</strong> value, HRB 69.6 4843.2 69.6 1<br />

Quantity Unit Estimated Standard Sensitivity Uncertainty<br />

error value uncertainty coefficient c<strong>on</strong>tributi<strong>on</strong><br />

Xi xi u(xi) ci ui(H) Preliminary test force, Fo N 0.020 0.011 0.080 0.001<br />

Total test force, F N 0.981 0.566 -0.075 -0.042<br />

Permanent depth of indentati<strong>on</strong>, h m 0.2 0.115 -0.500 -0.058<br />

Preliminary test force dwell time, to s 0.5 0.289 0.130 0.037<br />

Indentati<strong>on</strong> velocity, v mm/s -1<br />

0.3 0.173 0.000 0.000<br />

Total test force dwell time, t s 1 0.289 -0.017 -0.005<br />

Indenter ball diameter, d m 2 1.155 -0.012 -0.014<br />

Deformati<strong>on</strong> of frame, DF m 0.25 0.144 0.500 0.072<br />

Combined uncertainty, u c HRB Combined uncertainty, uc 0.109<br />

Expanded uncertainty, U e HRB Expanded uncertainty, Ue 0.219<br />

Uncertainty budget of hardness measurement at 70 HRB 70<br />

Quantity Unit Estimated Standard Sensitivity Uncertainty<br />

value uncertainty coefficient c<strong>on</strong>tributi<strong>on</strong><br />

X i x i u(x i) c i u i(H)<br />

<strong>Hardness</strong> machine, H phm HRB 69.6 0.109 1 0.109<br />

Readability,H res HRB 0 0.014 1 0.014<br />

Repeatability of measurement, H rep HRB 0 0.078 1 0.078<br />

Combined uncertainty, uc 0.135<br />

Measured hardness,H mh HRB 69.6 Expanded uncertainty, Ue 0.270<br />

Page 47 of 77

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!