Revisi<strong>on</strong>: <str<strong>on</strong>g>Final</str<strong>on</strong>g> <str<strong>on</strong>g>Report</str<strong>on</strong>g><br />

Date: May 25, 2012<br />

Annex A Data report forms<br />

<str<strong>on</strong>g>Report</str<strong>on</strong>g> Form 1 (For Rockwell hardness scale A)<br />

Metrology informati<strong>on</strong><br />

Model of hardness primary machine SHT-31<br />

Serial number of hardness primary machine 310039<br />

Manufacturer of hardness primary machine Akashi – Japan<br />

<strong>Hardness</strong> indicati<strong>on</strong> system Laser hologauge<br />

Test force system Lever type<br />

Relative error of preliminary test force +0.01 % (+0.001 kgf)<br />

Relative uncertainty (2) of preliminary test force 0,030 %<br />

Relative error of total test force +0.001 % (+0.0033 kgf)<br />

Relative uncertainty (2) of total test force 0.01%<br />

Uncertainty of depth measuring device 0.2 m<br />

Serial number of the indenter 53495<br />

The tip radius of curvature of the indenter 200m ± 4m<br />

The apex angle of c<strong>on</strong>e of the indenter 120 ± 0.1<br />

The calibrati<strong>on</strong> and measurement capability (CMC) 0.38 HRA<br />

Testing c<strong>on</strong>diti<strong>on</strong><br />

Velocity of the indenter when reaching the surface 80-200 m/s<br />

Holding time of preliminary test force 3 s<br />

Test force loading speed 33 s<br />

Total test force durati<strong>on</strong> time 4 s<br />

Time before reading permanent indentati<strong>on</strong> depth 4 s<br />

Envir<strong>on</strong>ment of comparis<strong>on</strong> measurement<br />

Measurement place <strong>Hardness</strong> lab. NIMT<br />

Measurement date 28-29 April 2010<br />

Envir<strong>on</strong>ment (23±1)C ,(50±10)%RH<br />

Deformati<strong>on</strong> of frame verificati<strong>on</strong><br />

Deformati<strong>on</strong> of frame (m)<br />

1 2 3<br />

0.14 0.14 0.15<br />