APMP.M.H-S3 Final Report Comparison on Hardness ... - BIPM
APMP.M.H-S3 Final Report Comparison on Hardness ... - BIPM
APMP.M.H-S3 Final Report Comparison on Hardness ... - BIPM
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
Revisi<strong>on</strong>: <str<strong>on</strong>g>Final</str<strong>on</strong>g> <str<strong>on</strong>g>Report</str<strong>on</strong>g><br />
Date: May 25, 2012<br />
<str<strong>on</strong>g>Report</str<strong>on</strong>g> Form 4 Uncertainty budget of measurement Page 6/6<br />
Uncertainty budget for 25 HRB<br />
Quantity Estimated Standard Sensitivity Uncertainty<br />
value<br />
uncertainty coefficient c<strong>on</strong>tributi<strong>on</strong><br />
X i x i<br />
xi<br />
uxi<br />
<br />
3<br />
C i u i (H )<br />
Preliminary test force, F 0 0.05 N 0.0289 0.1780 0.0051<br />
Total test force, F 0.5 N 0.2887 -0.1160 -0.0335<br />
Permanent depth of measuring device, h 0.1 µm 0.0577 0.5000 0.0289<br />
Preliminary test force dwell time, t 0 0.2 s 0.1155 -0.0617 -0.0071<br />
Indentati<strong>on</strong> velocity, v 10 µm/s 5.7735 0.0067 0.0387<br />
Total test force dwell time, t 0.2 s 0.1155 -0.0170 -0.0020<br />
Indenter diameter, d 0.1 µm 0.0577 0.0098 0.0006<br />
Deformati<strong>on</strong> of frame, DF 0.3 µm 0.1732 0.5000 0.0866<br />
Combined uncertainty, u c<br />
Expanded uncertainty, Ue<br />
0.1050<br />
k=2 0.21<br />
Page 35 of 77