APMP.M.H-S3 Final Report Comparison on Hardness ... - BIPM
APMP.M.H-S3 Final Report Comparison on Hardness ... - BIPM
APMP.M.H-S3 Final Report Comparison on Hardness ... - BIPM
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
Revisi<strong>on</strong>: <str<strong>on</strong>g>Final</str<strong>on</strong>g> <str<strong>on</strong>g>Report</str<strong>on</strong>g><br />
Date: May 25, 2012<br />
Data report forms Page 2/6<br />
<str<strong>on</strong>g>Report</str<strong>on</strong>g> Form 1 (For Rockwell hardness scale B)<br />
Metrology informati<strong>on</strong><br />
Model of hardness primary machine<br />
Serial number of hardness primary machine RNG10<br />
Manufacturer of hardness primary machine WAZAU<br />
<strong>Hardness</strong> indicati<strong>on</strong> system Laserinterferometer<br />
Test force system<br />
Relative error of preliminary test force<br />
Deth weight<br />
Relative uncertainty (2) of preliminary test force<br />
Relative error of total test force<br />
5*10^-4<br />
Relative uncertainty (2) of total test force 5*10^-4<br />
Uncertainty of depth measuring device 0.1 µm<br />
The calibrati<strong>on</strong> and measurement capability<br />
(CMC)<br />
0.5 HRB<br />
Testing c<strong>on</strong>diti<strong>on</strong><br />
Velocity of the indenter when reaching the surface 0.86 mm/s<br />
Holding time of preliminary test force 2.5 s<br />
Test force loading speed (final speed) 103 N/s<br />
Total test force durati<strong>on</strong> time 5.5 s<br />
Time before reading permanent indentati<strong>on</strong> depth 4.8 s<br />
Envir<strong>on</strong>ment of comparis<strong>on</strong> measurement<br />
Measurement place PTB<br />
Measurement date 1st June 2010<br />
Envir<strong>on</strong>ment (20 ± 1) °C<br />
Deformati<strong>on</strong> of frame verificati<strong>on</strong><br />
Deformati<strong>on</strong> of frame (m)<br />
1 2 3<br />
0.324 0.311 0.297<br />
Page 31 of 77