THE RELATIONSHIP BETWEEN CCT AND IOP READINGS Others report no linear correlation between CCT and IOP, pointing out that models are dependent on: population characteristics presence/absence of OHT number and type of surgical interventions ethnic groups pachymetric techniques depending on the corneal pachymetric method adopted, may not be able to confirm the true IOP of subjects tonometric techniques each can over or under estimate IOP
THE REICHERT OCULAR RESPONSE ANALYZER® (ORA) The ORA utilizes an air pulse to apply force to the cornea and an advanced electro-optical system to monitor the resultant corneal deformation. Alignment to the patient’s eye is fully automated. The difference between "inward" applanation pressure and "outward" applanation pressures defines corneal hysteresis, a measure of ocular rigidity. RORA measurement may not be possible in the presence of nystagmus
- Page 1 and 2: Tonometry Gillian J. McLellan BVMS
- Page 3 and 4: Glaucoma ------------ 25 mm Hg Nor
- Page 5 and 6: Mercury Column Manometry Courtesy C
- Page 7 and 8: Clinical issues impacting tonometri
- Page 9 and 10: Restraint vs vs. vs vs. . no restra
- Page 11 and 12: How to evaluate tonometry papers Ac
- Page 13 and 14: How to evaluate tonometry papers…
- Page 15 and 16: Schiøtz - Practical Use Topical a
- Page 17 and 18: ↓ Force ↑ Force Schiøtz - Prac
- Page 19 and 20: Maklakov Tonometer, First produced
- Page 21 and 22: Imbert Imbert-Fick Imbert Imbert-Fi
- Page 23 and 24: Goldmann applanation tonometry •A
- Page 25 and 26: Mackay Mackay-Marg Mackay Mackay-Ma
- Page 27 and 28: Tono Tono-pen Tono Tono-pen pen The
- Page 29 and 30: Clinical considerations for Mackay
- Page 31 and 32: Corneal Biomechanics • Applanatio
- Page 33 and 34: Rebound tonometer • Probe rebound
- Page 36 and 37: Non-contact device that flattens th
- Page 40 and 41: The Proview eye pressure monitor Ha
- Page 42 and 43: The Pascal DCT (Swiss Microtechnolo
THE REICHERT OCULAR RESPONSE ANALYZER® (ORA)<br />
The ORA utilizes an air pulse to apply<br />
force to the cornea and an advanced<br />
electro-optical system to monitor the<br />
resultant corneal deformation.<br />
Alignment to the patient’s eye is fully<br />
automated.<br />
The difference between "inward"<br />
applanation pressure and "outward"<br />
applanation pressures defines<br />
corneal hysteresis, a measure of<br />
ocular rigidity.<br />
RORA measurement may not be possible in the presence of nystagmus