GB USE OF THE HAND CONTROL Timer Pause on lower limit Pause on upper limit Load Speed Upper limit Lower limit Movement type Program number Comments: • When a program has been deleted, the display shows • The values shown in the ‘Display’ column are examples. They actually depend on the stored programs. Program table: 1 2 3 4 5 6 Page 16 7 8 9 10 11 PROGRAM 11 EMPTY 12 13 14 15 16
USE OF THE HAND CONTROL How to define the upper and lower movement limits • At the start of a session The MANUAL MODE is a way to set within the tolerance of a patient at the beginning of a session. Proceed as below: Beginning Switch the unit on To select the MANUAL MODE for upper limits by continually holding pressure on the key To set the pain level when reached, immediately press To select the manual mode for lower limits To set the pain level when reached, immediately press To start the session with the new movement limits Keys to press continuous press continuous press Page 17 Display Remarks KINETEC CENTURA V3.2 MOVEMENT VERIF. PLEASE WAIT MOVEMENT VERIF. MOTEoR: M1 M2 FLEX/EXT 50 STOP 15 100 FLEX/EXT 50 MANUaL 15 100 FLEX/EXT 50 MANUaL 150 150 FLEX/EXT 50 MANUaL 100 150 FLEX/EXT 25 MANUaL 25 150 FLEX/EXT 25 run 30 150 Check if the locking switch is in the following position: or The unit is moving to the upper limit of the movement. The new upper value limit of the movement is recorded. The unit is moving to the lower limit of the movement The new lower value limit of the movement is recorded. The angle display changes with current movement. Specific rules for synchronized movement: • You can only change the upper limit of the movement through the manual mode and only beyond the synchronization point. Comments: • The values shown in the ‘Display’ column are examples. They actually depend on the stored programs. GB
- Page 1 and 2: • Manuel d’utilisation • User
- Page 3 and 4: SUMMARY Page 1 Page DESCRIPTION 2-6
- Page 5 and 6: DESCRIPTION ELECTRICAL CONNECTION K
- Page 7 and 8: 30 32 20 31 DESCRIPTION 21 15 22 25
- Page 9 and 10: ASSEMBLY & TRANSPORT Base assembly
- Page 11 and 12: USE OF THE HAND CONTROL Reset time
- Page 13 and 14: USE OF THE HAND CONTROL How to set
- Page 15 and 16: USE OF THE HAND CONTROL Reading the
- Page 17: USE OF THE HAND CONTROL How to modi
- Page 21 and 22: INSTRUCTIONS FOR USE How to use the
- Page 23 and 24: INSTRUCTIONS FOR USE Patient set up
- Page 25 and 26: INSTRUCTIONS FOR USE Patient set up
- Page 27 and 28: INSTRUCTIONS FOR USE Patient set up
- Page 29 and 30: INSTRUCTIONS FOR USE Patient set up
- Page 31 and 32: INSTRUCTIONS FOR USE Patient set up
- Page 33 and 34: INSTRUCTIONS FOR USE Patient set up
- Page 35 and 36: PRODUCT INFORMATION MAINTENANCE Aft
- Page 37 and 38: Page 35 GB
USE OF THE HAND CONTROL<br />
How to define the upper and lower movement limits<br />
• At the start of a session<br />
The MANUAL MODE is a way to set within the tolerance of a patient at the beginning of a session.<br />
Proceed as below:<br />
Beginning<br />
Switch the unit on<br />
To select the MANUAL<br />
MODE for upper limits by<br />
continually holding<br />
pressure on the key<br />
To set the pain level when<br />
reached, immediately<br />
press<br />
To select the manual<br />
mode for lower limits<br />
To set the pain level when<br />
reached, immediately<br />
press<br />
To start the session with<br />
the new movement limits<br />
Keys to<br />
press<br />
continuous<br />
press<br />
continuous press<br />
Page 17<br />
Display Remarks<br />
KINETEC<br />
CENTURA V3.2<br />
MOVEMENT VERIF.<br />
PLEASE WAIT<br />
MOVEMENT VERIF.<br />
MOTEoR: M1 M2<br />
FLEX/EXT<br />
50 STOP 15 100<br />
FLEX/EXT<br />
50 MANUaL 15 100<br />
FLEX/EXT<br />
50 MANUaL 150 150<br />
FLEX/EXT<br />
50 MANUaL 100 150<br />
FLEX/EXT<br />
25 MANUaL 25 150<br />
FLEX/EXT<br />
25 run 30 150<br />
Check if the locking switch<br />
is in the following<br />
position: or<br />
The unit is moving to the<br />
upper limit of the<br />
movement.<br />
The new upper value limit<br />
of the movement is<br />
recorded.<br />
The unit is moving to the<br />
lower limit of the<br />
movement<br />
The new lower value limit<br />
of the movement is<br />
recorded.<br />
The angle display<br />
changes with current<br />
movement.<br />
Specific rules for synchronized movement:<br />
• You can only change the upper limit of the movement through the manual mode and only beyond the<br />
synchronization point.<br />
Comments:<br />
• The values shown in the ‘Display’ column are examples. They actually depend on the stored<br />
programs.<br />
GB