A Understanding the ChromAlign Process<br />

Aligning Over Different Abundance Levels<br />

Aligning Over Different Abundance Levels<br />

Biomarker discovery detects low abundance components in the presence of high abundance<br />

components that remain essentially unchanged between the controls and samples. The<br />

following example shows the alignment of low abundance components in the presence of high<br />

abundance components. Figure 20 shows reconstructed ion chromatograms of a horse<br />

myoglobin peptide added to the nine-protein mix at three different concentrations (500 fmol<br />

[red], 250 fmol [green], and 100 fmol [blue]) and compared to the control sample (1000 fmol<br />

[black]).<br />

Figure 20. Original reconstructed ion chromatograms<br />

Abundance<br />

0 50000 100000 150000<br />

49 50 51<br />

Time (Min)<br />

52 53<br />

32 <strong>SIEVE</strong> <strong>User</strong> <strong>Guide</strong> Thermo Fisher