Data Interfaces<br />

3 You<br />

Connection Options<br />

may need to use an external power supply to operate peripheral devices.<br />

Preparation See chapter “Getting Started,“ sections “COM1 Pin Assignment Chart, “COM2 Pin<br />

Assignment Chart,“ “PS2 Pin Assignment Chart“ for pin assignments and cabling<br />

diagrams.<br />

Printer Connection Options The following printers can be connected to the standard COM1 and COM2<br />

interfaces or the optional UniCOM universal interface:<br />

– YDP20 (user-defined interface parameters)<br />

– YDP14IS (strip or label printer)<br />

– YDP04IS (strip or label printer)<br />

– Universal printer (user-definable transmission parameters)<br />

Device Connection Options In addition, the following devices can be connected to the standard COM1 and<br />

COM2 interfaces:<br />

– Foot / Hand switch, COM1 only<br />

– Second printer<br />

– Remote display<br />

– PC (RS-232 interface)<br />

– Third weighing platform (RS-232 interface)<br />

– External checkweighing display (stop light) via a digital I/O (<strong>Sartorius</strong> standard),<br />

COM1 only<br />

– To PS2: Barcode scanner/External keyboard<br />

The following devices can be connected to the optional UniCOM universal interface:<br />

– WP3 RS-232/RS-485<br />

– PC (RS-232 interface)<br />

– Second printer (external power source required)<br />

– Remote display<br />

– Digital I/O<br />

– Current interface (0/4...20 mA), voltage (0...10 V)<br />

– PLC with Profibus DP<br />

– Ethernet<br />

Connection Options<br />

for other Weighing Platforms The <strong>Combics</strong> 3 model enables you to connect a 2nd or 3rd weighing platform.<br />

You can use both the COM1, COM2 or the UniCOM port.<br />

The standard COM1 and COM2 port is operated in RS-232 mode. The following<br />

modes are available for a 3rd weighing platform:<br />

– SBI<br />

– IS-232 (factory setting)<br />

– ADC-232<br />

Operating Modes UniCOM interfaces can be operated in either the RS-232 or RS-485 mode.<br />

The third weighing platform can be operated in any of the following modes:<br />

– SBI (RS-232 mode)<br />

– IS-232 (RS-232 mode)<br />

– ADC-232 (RS-232 mode)<br />

– IS-485 (RS-485 mode, xBPI mode, factory setting)<br />

– ADC-485 (RS-485 mode)<br />

The standard COM1 and COM2 ports or the optional universal UniCOM interface<br />

can be used as a printer interface.<br />

98 Operating Instructions <strong>Combics</strong> <strong>Complete</strong> <strong>Scale</strong>s