Data Interfaces<br />
3 You<br />
Connection Options<br />
may need to use an external power supply to operate peripheral devices.<br />
Preparation See chapter “Getting Started,“ sections “COM1 Pin Assignment Chart, “COM2 Pin<br />
Assignment Chart,“ “PS2 Pin Assignment Chart“ for pin assignments and cabling<br />
diagrams.<br />
Printer Connection Options The following printers can be connected to the standard COM1 and COM2<br />
interfaces or the optional UniCOM universal interface:<br />
– YDP20 (user-defined interface parameters)<br />
– YDP14IS (strip or label printer)<br />
– YDP04IS (strip or label printer)<br />
– Universal printer (user-definable transmission parameters)<br />
Device Connection Options In addition, the following devices can be connected to the standard COM1 and<br />
COM2 interfaces:<br />
– Foot / Hand switch, COM1 only<br />
– Second printer<br />
– Remote display<br />
– PC (RS-232 interface)<br />
– Third weighing platform (RS-232 interface)<br />
– External checkweighing display (stop light) via a digital I/O (<strong>Sartorius</strong> standard),<br />
COM1 only<br />
– To PS2: Barcode scanner/External keyboard<br />
The following devices can be connected to the optional UniCOM universal interface:<br />
– WP3 RS-232/RS-485<br />
– PC (RS-232 interface)<br />
– Second printer (external power source required)<br />
– Remote display<br />
– Digital I/O<br />
– Current interface (0/4...20 mA), voltage (0...10 V)<br />
– PLC with Profibus DP<br />
– Ethernet<br />
Connection Options<br />
for other Weighing Platforms The <strong>Combics</strong> 3 model enables you to connect a 2nd or 3rd weighing platform.<br />
You can use both the COM1, COM2 or the UniCOM port.<br />
The standard COM1 and COM2 port is operated in RS-232 mode. The following<br />
modes are available for a 3rd weighing platform:<br />
– SBI<br />
– IS-232 (factory setting)<br />
– ADC-232<br />
Operating Modes UniCOM interfaces can be operated in either the RS-232 or RS-485 mode.<br />
The third weighing platform can be operated in any of the following modes:<br />
– SBI (RS-232 mode)<br />
– IS-232 (RS-232 mode)<br />
– ADC-232 (RS-232 mode)<br />
– IS-485 (RS-485 mode, xBPI mode, factory setting)<br />
– ADC-485 (RS-485 mode)<br />
The standard COM1 and COM2 ports or the optional universal UniCOM interface<br />
can be used as a printer interface.<br />
98 Operating Instructions <strong>Combics</strong> <strong>Complete</strong> <strong>Scale</strong>s