Data Interfaces<br />

Data Interfaces<br />

The indicator is equipped with the following data interfaces:<br />

– COM1: Standard data interface (RS-232)<br />

– COM2: Standard data interface (RS-232)<br />

– UniCOM: Universal data interface (optional)<br />

Interfaces can be configured in the Setup menu for different input and output<br />

functions (e.g. printer, 3rd weighing platform, PC, checkweighing).<br />

The optional UniCOM interface can be used as an RS-232, RS-485/RS-422 or<br />

analog output (voltage/current interface, galvanically separated digital I/Os,<br />

profibus, Ethernet), see also “Accessories.“<br />

A barcode scanner or a keyboard can be connected via the PS/2 -socket or the<br />

corresponding screw terminals (IP69K).<br />

Features – IP44 indicator:<br />

Connect via a 25-pin D-Sub female connector<br />

– IP69K indicator:<br />

Route the connecting cable from the peripheral device to the indicator via<br />

a cable gland. The free cable ends are connected via the screw terminals.<br />

3 Warning<br />

96 Operating Instructions <strong>Combics</strong> <strong>Complete</strong> <strong>Scale</strong>s<br />

when using third-party RS-232 connecting cables:<br />

The pin assignments may not be compatible with <strong>Sartorius</strong> equipment.<br />

Check the pin assignment against the cabling diagrams and disconnect any lines<br />

that are not assigned. Failure to do so may cause malfunction, damage or even<br />

completely ruin your indicator and/or peripheral device(s).