MAN-Combics Complete Scale CAW3P CAW3S CAH3-e - Sartorius
MAN-Combics Complete Scale CAW3P CAW3S CAH3-e - Sartorius
MAN-Combics Complete Scale CAW3P CAW3S CAH3-e - Sartorius
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
Operation<br />
SQmin Operation<br />
Configuration in Service Mode The SQmin value can only be entered in Service mode.<br />
t Remove the cap.<br />
t Slide the menu access switch to the left (= “open“ position).<br />
If the device is part of a verified weighing facility, this will only be possible if the<br />
verification seal is broken. The weighing equipment must then be verified again.<br />
t Activate the Service mode, see “Service Mode.“<br />
t In the “Device parameters“ menu, select “SQmin“ and open<br />
using the “O“ soft key.<br />
t Select “SQmin input“ using the “O“ soft key.<br />
t Select “SQmin WP1“ using the “O“ soft key, enter the value via the keypad<br />
(in this example 0.100 kg) and save using the “l“ soft key or cancel using “ESC.“<br />
t Then enter the values for WP 2 and WP 3.<br />
t Press M or “oo“ to exit the Setup menu.<br />
t Slide the menu access switch to the right (= “closed“ position) and reattach the<br />
cap.<br />
y The device is now in normal weighing mode.<br />
Configuration in Weighing Mode t In the “Device parameters“ menu, select “SQmin“ and open<br />
using the “O“ soft key.<br />
t Select “Display.“.<br />
t Open using the “O“ soft key.<br />
t Confirm the factory setting “oYes“ using the “l“ soft key.<br />
t Press the “o“ soft key several times to switch to the next highest menu.<br />
92 Operating Instructions <strong>Combics</strong> <strong>Complete</strong> <strong>Scale</strong>s<br />
The SQmin display must be turned on to use the SQmin function.<br />
t In the “Fn key:SQmin display:“ menu, select using the “O“ soft<br />
key and confirm using the “l“ soft key.<br />
t Press M or “oo“ to exit the Setup menu.