Operation<br />
C.INTER<br />
C.Inter<br />
C.EXT.DEF.<br />
CI<br />
Just.-Intern<br />
C.EXT.DEF.<br />
Internal adjustment for CAH Models...<br />
System Requirements – The menu item “Intern. cal./adjust.“ is preset for scales with<br />
internal adjustment.<br />
– The scale housing has a built-in motorized adjustment weight.<br />
– The calibration/adjustment sequence is preset to manual adjustment in the<br />
“Calibration/adjustment:Cal/adj. sequence:<br />
Cal. with manual adjust.,“ see section “Function Allocation<br />
of the J Key for Calibration/Adjustment“<br />
90 Operating Instructions <strong>Combics</strong> <strong>Complete</strong> <strong>Scale</strong>s<br />
Example: Internal calibration and then adjustment (weighing parameters: factory settings<br />
for CAH... models)<br />
Procedure t Press the ( key to unload and zero the scale.<br />
t Press J to start the internal calibration.<br />
y The display “C.INTER“ appears while the internal adjustment weight is placed<br />
on the platform.<br />
y Then the display “CI“ appears.<br />
y After a brief pause, the difference since the last span adjustment is displayed.<br />
t Press J to start internal adjustment or press ( to cancel the calibration/<br />
adjustment.<br />
y When calibration/adjustment is complete, the GMP-compliant printout is<br />
generated. If the adjustment procedure is canceled (only calibration is<br />
performed), the last two lines, “Internal adjustment” and “Diff. + 0.000 kg”<br />
are not printed.<br />
y The device is now in normal weighing mode.