External calibration<br />

Nom. + 10000 g<br />

Diff. + 1 g<br />

--------------------<br />

24.10.2010 10:15<br />

Type CAIS3<br />

Vers.no. 1.02.101110<br />

BVers. 01-63-02<br />

--------------------<br />

External calibration<br />

Nom. + 10000 g<br />

Diff. + 1 g<br />

External adjustment<br />

Diff. + 0 g<br />

--------------------<br />

24.10.2010 10:15<br />

Name:<br />

--------------------<br />

t Place the calibration/adjustment weight on the weighing platform.<br />

y The difference between the measured value and the true weight of the sample<br />

will be displayed with plus/minus signs.<br />

y A printout will be generated if the adjustment is not carried out and the<br />

procedure is stopped by pressing the ( key.<br />

t Start adjustment via the J key (cancel calibration/adjustment via ().<br />

y The adjustment weight is displayed once adjustment is finished.<br />

y A GMP-compliant printout is generated.<br />

Operation<br />

Operating Instructions <strong>Combics</strong> <strong>Complete</strong> <strong>Scale</strong>s 89