Automatic Taring<br />

The first weight on the scale that exceeds the preset minimum capacity is stored in<br />

the tare memory at stability. The values for subsequent loads are stored as weight<br />

values.<br />

The scale returns to the initial state when the load on the scale is less than 50% of<br />

the minimum load.<br />

Operating Menu Setting: “Application parameters: Autotare 1st<br />

weight:Yes“<br />

Minimum load for automatic taring and automatic printing<br />

To tare container weights automatically, you need to set a minimum load in the<br />

Setup menu, under:<br />

“Application parameters:Min. load f. auto.<br />

taring/printout:o 10 digits“<br />

You can set the following for the minimum load:<br />

1 digit (no minimum load)<br />

2 digits<br />

5 digits<br />

10 digits<br />

20 digits<br />

50 digits<br />

100 digits<br />

200 digits<br />

500 digits<br />

1000 digits<br />

The “digits“ here refer to the scale intervals for the connected weighing platform.<br />

If the interval is 1 g and 1000 digits are required, the minimum load is 1000 g<br />

(1000 intervals).<br />

If the weighing platform interval is 5 g and the same number of digits as above are<br />

required, the minimum load is 5000 g.<br />

Once the minimum load has been exceeded, the weighing platform is tared<br />

automatically and/or an automatic printout is generated.<br />

Automatic Printing<br />

The first weight value that exceeds the minimum load is printed. If the menu item<br />

is also activated for automatic taring, it is only tared when the minimum load is<br />

exceeded. In this case, an automatic printout would only be generated when the<br />

second weight value exceeds the minimum load.<br />

Setting in the Setup Menu: “Device parameters:Config. printout:Once at<br />

stability:On“<br />

Weighing using up to three weighing platforms<br />

Press the n key to toggle between three weighing platforms.<br />