MAN-Combics Complete Scale CAW3P CAW3S CAH3-e - Sartorius
MAN-Combics Complete Scale CAW3P CAW3S CAH3-e - Sartorius
MAN-Combics Complete Scale CAW3P CAW3S CAH3-e - Sartorius
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
Automatic Taring<br />
The first weight on the scale that exceeds the preset minimum capacity is stored in<br />
the tare memory at stability. The values for subsequent loads are stored as weight<br />
values.<br />
The scale returns to the initial state when the load on the scale is less than 50% of<br />
the minimum load.<br />
Operating Menu Setting: “Application parameters: Autotare 1st<br />
weight:Yes“<br />
Minimum load for automatic taring and automatic printing<br />
To tare container weights automatically, you need to set a minimum load in the<br />
Setup menu, under:<br />
“Application parameters:Min. load f. auto.<br />
taring/printout:o 10 digits“<br />
You can set the following for the minimum load:<br />
1 digit (no minimum load)<br />
2 digits<br />
5 digits<br />
10 digits<br />
20 digits<br />
50 digits<br />
100 digits<br />
200 digits<br />
500 digits<br />
1000 digits<br />
The “digits“ here refer to the scale intervals for the connected weighing platform.<br />
If the interval is 1 g and 1000 digits are required, the minimum load is 1000 g<br />
(1000 intervals).<br />
If the weighing platform interval is 5 g and the same number of digits as above are<br />
required, the minimum load is 5000 g.<br />
Once the minimum load has been exceeded, the weighing platform is tared<br />
automatically and/or an automatic printout is generated.<br />
Automatic Printing<br />
The first weight value that exceeds the minimum load is printed. If the menu item<br />
is also activated for automatic taring, it is only tared when the minimum load is<br />
exceeded. In this case, an automatic printout would only be generated when the<br />
second weight value exceeds the minimum load.<br />
Setting in the Setup Menu: “Device parameters:Config. printout:Once at<br />
stability:On“<br />
Weighing using up to three weighing platforms<br />
Press the n key to toggle between three weighing platforms.<br />
Operation<br />
Operating Instructions <strong>Combics</strong> <strong>Complete</strong> <strong>Scale</strong>s 77