Operating Design<br />

Operating<br />

Design<br />

You can use the <strong>Combics</strong> 3 to<br />

record weight values from 1 to<br />

3 weighing platforms, calculate<br />

and display weight values through<br />

application programs, and assign<br />

IDs to the samples weighed.<br />

Configure the indicator first, using<br />

the Setup menu for the desired<br />

application program (printer<br />

settings, etc.). Then you can begin<br />

weighing.<br />

When a key is pressed that does<br />

not have an active operating mode<br />

function, an acoustical signal<br />

(double beep) sounds and the<br />

message “——-—“ is displayed for<br />

2 seconds. The display then<br />

returns to the previous screen<br />

content.<br />

52 Operating Instructions <strong>Combics</strong> <strong>Complete</strong> <strong>Scale</strong>s<br />

Weighing Mode Operating Design<br />

Labeled Keys<br />

Some keys have a second function, activated by pressing and holding the key for<br />

over 2 seconds. Whether a function is available depends on the indicator operating<br />

state and operating menu settings.<br />

e On/Off key<br />

y standby is displayed in Standby mode.<br />

n If a second weighing platforms is connected, this key toggles the display between<br />

the two readouts.<br />

( – Press briefly: Zero<br />

– Press briefly: cancels calibration/adjustment<br />

– Press longer than 2 seconds: displays the adjustment/configuration counter<br />

) Tare the scale: press briefly.<br />

k Displays the second weight unit or SQmin (depending on the settings, see<br />

“Operation,“ “SQmin Function“).<br />

J Starts calibration or adjustment.<br />

p – For printing:<br />

Press briefly.<br />

– Prints GMP footer:<br />

Press longer than 2 seconds.<br />

K Toggles unit between normal and 10-fold higher resolution.<br />

L Toggles the display between<br />

gross value (net value plus tare) and<br />

net value (gross value minus tare).<br />

c – Quits an application or deletes an input character:<br />

Press the key less than 2 seconds.<br />

– Deletes the entire input string:<br />

Press the key longer than 2 seconds.