Configuring Weighing Platforms<br />

--------------------<br />

14.01.2010 13:50<br />

Type <strong>CAW3P</strong>1-6ED-LCE<br />

Ser.no. 12345678<br />

Vers.no. 1.02.101110<br />

BVers. 01-63-02<br />

--------------------<br />

Set preload<br />

completed<br />

--------------------<br />

14.01.2010 13:52<br />

Name:<br />

--------------------<br />

Set preload<br />

Setup Information – Setting the preload when weighing in legal metrology is only possible when the<br />

menu access switch is open.<br />

– The J key must be assigned the “Set preload“ function, see “Function<br />

Allocation of the J Key for Linearization and Setting/Deleting the Preload.“<br />

– Activating the display of geographical data has no effect on this function.<br />

3 Once<br />

50 Operating Instructions <strong>Combics</strong> <strong>Complete</strong> <strong>Scale</strong>s<br />

the preload has been set, the J key must be reallocated back to its original function<br />

in the Setup menu, e.g. external calibration/adjustment with default weights.<br />

Procedure t For scales used in legal metrology, slide the menu access switch to the left (= “open“<br />

position).<br />

t Press e turn the device back on.<br />

y The <strong>Sartorius</strong> logo is displayed briefly, after which the device is in normal weighing<br />

mode.<br />

t Press ( to unload and zero the scale.<br />

y Display after the scale has been zeroed.<br />

t Place the preload weight on the weighing platform.<br />

t Press J to start “Set preload.“<br />

y The display “SET PREL“ appears briefly.<br />

After a short period of time the preload will be applied and the indicator will<br />

automatically switch back to weighing mode.<br />

After the “Set Preload” operation has been completed, the scale is zeroed.<br />

When the “Set preload“ function is complete, the GMP-compliant printout shown on<br />

the left is generated.<br />

t Slide the menu access switch to the right (= “closed“ position).