External Linearization with the Factory-Defined<br />
Weight (Default Weights)<br />
Setup Information – This function is accessible only if the software and the functionality of the<br />
connected weighing platform permit this operation.<br />
– External linearization when weighing in legal metrology is only possible when<br />
the menu access switch is open.<br />
– The J key must be assigned the “External linearization“ function, see<br />
“Function Allocation of the J Key for Linearization and Setting/Deleting the<br />
Preload.“<br />
– Activating the display of geographical data has no effect on this function.<br />
3 Once<br />
Configuring Weighing Platforms<br />
linearization has been completed, the J key must be reallocated back to<br />
its original function in the Setup menu, e.g. external calibration/adjustment with<br />
default weights.<br />
Procedure t For scales used in legal metrology, slide the menu access switch to the left<br />
(= “open“ position).<br />
y The display goes out and the device restarts. Then weighing mode is active.<br />
t Press ( to unload and zero the scale.<br />
t Press J to start the external linearization.<br />
y The display “L.EXT.D“ appears briefly.<br />
Additional steps are explained in “External Linearization with User-Defined<br />
Weights.“<br />
Operating Instructions <strong>Combics</strong> <strong>Complete</strong> <strong>Scale</strong>s 47