Configuring Weighing Platforms<br />

Function Allocation of the J Key for Linearization and Setting/<br />

Deleting the Preload<br />

Purpose The J key is normally used for the calibration/adjustment function.<br />

The following additional functions can be allocated to the key when the Service<br />

mode is activated:<br />

– External linearization with default weights<br />

– External linearization with entered linearization weights<br />

– Internal linearization (for external IS platforms only)<br />

– Set preload<br />

– Delete preload<br />

3 Once<br />

46 Operating Instructions <strong>Combics</strong> <strong>Complete</strong> <strong>Scale</strong>s<br />

linearization has been completed, or after a preload has been set or deleted<br />

the function of the J key must be reallocated back to its original function in the<br />

Setup menu, e.g. external calibration/adjustment with default weights.<br />

Procedure t Remove the cap.<br />

t Slide the menu access switch to the left (= “open“ position).<br />

t Activate the Service mode, see “Service Mode.“<br />

t Select weighing platform “WP 1“ in the “Device Parameters“<br />

menu item.<br />

t If the “Internal“ setting is not already activated (marked by “o“), select<br />

the setting using the “Q“ or “q“ soft key and confirm with “O.“<br />

y The menu for the “WP-1 INTERNAL“ device parameters is displayed.<br />

t Use the “Q“ or “q“ soft key to select and open the “Calibration/<br />

adjustment“ menu using “O.“<br />

y The “CAL key function“ submenu is displayed.<br />

t Use the “Q“ or “q“ soft key to select the corresponding menu item and confirm<br />

with “l.“<br />

y The menu item is marked by a circle “o.“<br />

Note<br />

The functions that can be configured in the “CAL key function“<br />

submenu depends on the selected weighing platform and its configuration data.<br />

Functions that cannot be accessed are not displayed in the selection screen.<br />

Note<br />

The function set in the “CAL key function“ menu is carried out in the<br />

normal weighing mode because when you exit the Setup menu the Service mode is<br />

deactivated. To perform the function on a digital weighing platform (such as an IS<br />

platform), however, it must be carried out in Service mode.<br />

t The procedure is as follows: after selecting the desired function and exiting<br />

the Setup menu, reactivate Service mode again and then exit the Setup menu<br />

immediately by pressing the M key or the “oo“ soft key.<br />

y The scale is now in Service mode without this being displayed.<br />

t Trigger the previously set function using the J key.<br />

y The display shows “S_CAl:“ indicating that the scale is in Service mode.<br />

t The corresponding menu item can be selected using the “Select“ soft key<br />

and carried out using the J key.<br />

t If you cancel the function using the ( key or by restarting the scale using<br />

the e key, you will exit the Service mode.