Configuring Weighing Platforms<br />
Function Allocation of the J Key for Linearization and Setting/<br />
Deleting the Preload<br />
Purpose The J key is normally used for the calibration/adjustment function.<br />
The following additional functions can be allocated to the key when the Service<br />
mode is activated:<br />
– External linearization with default weights<br />
– External linearization with entered linearization weights<br />
– Internal linearization (for external IS platforms only)<br />
– Set preload<br />
– Delete preload<br />
3 Once<br />
46 Operating Instructions <strong>Combics</strong> <strong>Complete</strong> <strong>Scale</strong>s<br />
linearization has been completed, or after a preload has been set or deleted<br />
the function of the J key must be reallocated back to its original function in the<br />
Setup menu, e.g. external calibration/adjustment with default weights.<br />
Procedure t Remove the cap.<br />
t Slide the menu access switch to the left (= “open“ position).<br />
t Activate the Service mode, see “Service Mode.“<br />
t Select weighing platform “WP 1“ in the “Device Parameters“<br />
menu item.<br />
t If the “Internal“ setting is not already activated (marked by “o“), select<br />
the setting using the “Q“ or “q“ soft key and confirm with “O.“<br />
y The menu for the “WP-1 INTERNAL“ device parameters is displayed.<br />
t Use the “Q“ or “q“ soft key to select and open the “Calibration/<br />
adjustment“ menu using “O.“<br />
y The “CAL key function“ submenu is displayed.<br />
t Use the “Q“ or “q“ soft key to select the corresponding menu item and confirm<br />
with “l.“<br />
y The menu item is marked by a circle “o.“<br />
Note<br />
The functions that can be configured in the “CAL key function“<br />
submenu depends on the selected weighing platform and its configuration data.<br />
Functions that cannot be accessed are not displayed in the selection screen.<br />
Note<br />
The function set in the “CAL key function“ menu is carried out in the<br />
normal weighing mode because when you exit the Setup menu the Service mode is<br />
deactivated. To perform the function on a digital weighing platform (such as an IS<br />
platform), however, it must be carried out in Service mode.<br />
t The procedure is as follows: after selecting the desired function and exiting<br />
the Setup menu, reactivate Service mode again and then exit the Setup menu<br />
immediately by pressing the M key or the “oo“ soft key.<br />
y The scale is now in Service mode without this being displayed.<br />
t Trigger the previously set function using the J key.<br />
y The display shows “S_CAl:“ indicating that the scale is in Service mode.<br />
t The corresponding menu item can be selected using the “Select“ soft key<br />
and carried out using the J key.<br />
t If you cancel the function using the ( key or by restarting the scale using<br />
the e key, you will exit the Service mode.