If the weighing platform has multiple load cells, multiply the nominal capacity<br />

accordingly.<br />

Example:<br />

The weighing platform has 4 load cells each at 50 kg.<br />

The nominal capacity is 4 x 50 kg = 200 kg.<br />

In the example shown here, the weighing platform consists of one load cell with a<br />

maximum capacity of 10 kg.<br />

t Press 10.00 and confirm with the “l“ soft key.<br />

y The “Resolution“ input field is selected.<br />

The smallest digit “d“ is entered in this field in “kg“. The value must correspond<br />

to the “D“ entry in “ADC configuration: Standard:<br />

Ranges: Single-range mode.“<br />

In the example, this should be d = 0.002 kg.<br />

t Press 0.002 and confirm with the “l“ soft key.<br />

y The “Sensitivity 1“ input field is selected.<br />

If the weighing platform has multiple load cells, enter the sensitivity either<br />

– under “Sensitivity 1“ … “Sensitivity 4“ as an individual<br />

value or<br />

– under “Sensitivity 1“ as an average value.<br />

If an average value is entered for all cells, or if fewer than 4 load cells are<br />

present, enter “0“ in the remaining fields. Range of permitted values: 0.01…5<br />

mV/V.<br />

t Press 1.944, e.g. to enter a sensitivity of 1.944 mV/V.<br />

t Confirm using the “l“ soft key.<br />

y The “Sensitivity 2“ input field is selected.<br />

t Either enter a value or set all remaining input fields to “0.“<br />

t Confirm each using the “l“ soft key.<br />

t Press the “o“ soft key to exit the menu.<br />

Configuring Weighing Platforms<br />

t Use the “q“ soft key to select the “Save parameters“ submenu.<br />

t To save the configuration, select “Yes“ and confirm using the l soft key.<br />

In the first line of the display, the message “Data stored“ is shown<br />

briefly.<br />

Then the program then returns to the “No“ display status.<br />

t Use the “o“ to return to the next menu level.<br />

t Press M or “oo“ to exit the Setup menu.<br />

t Slide the menu access switch to the right (= “closed“ position) and reattach the<br />

cap.<br />

y The display goes out and the device restarts. Then weighing mode is active.<br />