MAN-Combics Complete Scale CAW3P CAW3S CAH3-e - Sartorius
MAN-Combics Complete Scale CAW3P CAW3S CAH3-e - Sartorius
MAN-Combics Complete Scale CAW3P CAW3S CAH3-e - Sartorius
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
C.EXT.D<br />
C.EXT.DEF.<br />
Configuring Weighing Platforms<br />
40 Operating Instructions <strong>Combics</strong> <strong>Complete</strong> <strong>Scale</strong>s<br />
t Press e to turn off the device.<br />
t Press e turn the device back on.<br />
y The <strong>Sartorius</strong> logo is displayed briefly, after which the device is in normal<br />
weighing mode.<br />
t Press ( to unload and zero the scale.<br />
t Press J to start the external adjustment.<br />
y The display “C.EXT.D“ appears briefly.<br />
Note<br />
If the display of geographical data (elevation and latitude or gravity) is activated,<br />
this data is displayed and confirmed each with a press of the J key (you can<br />
cancel the calibration / adjustment process using the ( key).<br />
See also “Entering Geographical Data for Use in Legal Metrology“ in this chapter.<br />
Procedure y The target value of the required adjustment weight (5.000 kg in the example) is<br />
displayed as a negative value.<br />
t Place the required adjustment weight on the platform.<br />
Note<br />
If the calibration/adjustment sequence is set to automatic (“Calibration/<br />
adjustment:Cal./adj. sequence:Cal., then auto<br />
adjust“ menu, see “Function Allocation of the allocation of the J Key for<br />
Calibration/Adjustment“) and the adjustment weight consists of several pieces, then<br />
they should be placed on the platform in series at short intervals.<br />
When the weighing instrument has stabilized, the weight on the scale is accepted<br />
as the calibration weight after a predefined interval, and the weighing instrument is<br />
calibrated/adjusted with this weight.<br />
The difference since the most recent span adjustment is not displayed; this value is<br />
output only on GMP-compliant printouts.<br />
y After a brief pause, the difference since the last span adjustment is displayed<br />
(calibration).<br />
Note<br />
This display only appears for the “Cal. then manual adj.“<br />
setting (see previous note). If “Cal. then auto adj.“ is active, the<br />
calibration/adjustment procedure cannot be cancelled.<br />
To stop the procedure after calibration and before adjustment takes place, press the<br />
( key (only if “Cal. then manual adj.“ is active).<br />
t Press J to adjust the scale (only if “Cal. then manual adj.“<br />
is active).<br />
y At the conclusion of the calibration procedure, the calibration weight is<br />
displayed as a positive value.