C.EXT.D<br />

C.EXT.DEF.<br />

Configuring Weighing Platforms<br />

40 Operating Instructions <strong>Combics</strong> <strong>Complete</strong> <strong>Scale</strong>s<br />

t Press e to turn off the device.<br />

t Press e turn the device back on.<br />

y The <strong>Sartorius</strong> logo is displayed briefly, after which the device is in normal<br />

weighing mode.<br />

t Press ( to unload and zero the scale.<br />

t Press J to start the external adjustment.<br />

y The display “C.EXT.D“ appears briefly.<br />

Note<br />

If the display of geographical data (elevation and latitude or gravity) is activated,<br />

this data is displayed and confirmed each with a press of the J key (you can<br />

cancel the calibration / adjustment process using the ( key).<br />

See also “Entering Geographical Data for Use in Legal Metrology“ in this chapter.<br />

Procedure y The target value of the required adjustment weight (5.000 kg in the example) is<br />

displayed as a negative value.<br />

t Place the required adjustment weight on the platform.<br />

Note<br />

If the calibration/adjustment sequence is set to automatic (“Calibration/<br />

adjustment:Cal./adj. sequence:Cal., then auto<br />

adjust“ menu, see “Function Allocation of the allocation of the J Key for<br />

Calibration/Adjustment“) and the adjustment weight consists of several pieces, then<br />

they should be placed on the platform in series at short intervals.<br />

When the weighing instrument has stabilized, the weight on the scale is accepted<br />

as the calibration weight after a predefined interval, and the weighing instrument is<br />

calibrated/adjusted with this weight.<br />

The difference since the most recent span adjustment is not displayed; this value is<br />

output only on GMP-compliant printouts.<br />

y After a brief pause, the difference since the last span adjustment is displayed<br />

(calibration).<br />

Note<br />

This display only appears for the “Cal. then manual adj.“<br />

setting (see previous note). If “Cal. then auto adj.“ is active, the<br />

calibration/adjustment procedure cannot be cancelled.<br />

To stop the procedure after calibration and before adjustment takes place, press the<br />

( key (only if “Cal. then manual adj.“ is active).<br />

t Press J to adjust the scale (only if “Cal. then manual adj.“<br />

is active).<br />

y At the conclusion of the calibration procedure, the calibration weight is<br />

displayed as a positive value.