C.EXT.D<br />

C.EXT.DEF.<br />

t Use the “O“ soft key to open the CAL key function“ menu.<br />

y The “CAL key function“ submenu is displayed.<br />

t Use the “Q“ or “q“ soft key to select the corresponding menu item and confirm<br />

with “l.“<br />

y The menu item is marked by a circle “o.“<br />

Note<br />

The functions that can be configured in the “CAL key function“<br />

submenu depends on the selected weighing platform and its configuration data.<br />

Functions that cannot be accessed are not displayed in the selection screen.<br />

Important Note!<br />

The function set in the “CAL key function“ menu is carried out in the<br />

normal weighing mode because when you exit the Setup menu the Service mode is<br />

deactivated. To perform the function on a digital weighing platform (such as an<br />

IS platform), however, it must be carried out in Service mode.<br />

t The procedure is as follows: after selecting the desired function and exiting<br />

the Setup menu, reactivate Service mode again and then exit the Setup menu<br />

immediately by pressing the M key or the “oo“ soft key.<br />

y The scale is now in Service mode without this being displayed.<br />

t Trigger the previously set function using the J key.<br />

Configuring Weighing Platforms<br />

12 y The display shows “S_CAl:“ indicating that the scale is in Service mode.<br />

t The corresponding menu item can be selected using the “Select“ soft key<br />

and carried out using the J key.<br />

t If you cancel the function using the ( key or by restarting the scale using<br />

the e key, you will exit the Service mode.<br />

External Calibration/Adjustment<br />

with Factory-Defined Weight (Default Weight)<br />

Configuration If not already selected (factory setting, marked by “o“ when active), select the<br />

“CAL key function:Ext. cal./adj.; factorydef.<br />

wt.“ menu item (external adjustment with a factory-defined, standard<br />

weight).<br />

t Use the “Q“ or “q“ soft key to select this menu item and confirm with “l.“<br />

y The menu item is marked by a circle “o.“<br />

Note:<br />

The menu items “Ext. cal./adj.; factory-def. wt.“<br />

(external adjustment with a factory-defined, standard weight), “Ext. cal./<br />

adj.; user-defi ned wt.“ (external adjustment with<br />

a user-defined weight) and “Key blocked“ can also be accessed without<br />

activating the Service mode.<br />