Configuring Weighing Platforms<br />
38 Operating Instructions <strong>Combics</strong> <strong>Complete</strong> <strong>Scale</strong>s<br />
Function Allocation of the allocation<br />
for the J Key for Calibration/Adjustment<br />
Purpose The J key is used for the calibration/adjustment function. Key settings can be<br />
changed when the Service mode is activated:<br />
Procedure t Remove the cap.<br />
t Slide the menu access switch to the left (= “open“ position).<br />
t Activate the Service mode, see “Service Mode.“<br />
t Select weighing platform “WP 1“ in the “Device Parameters“<br />
menu item.<br />
t If the “Internal“ setting is not already activated (marked by “o“), select<br />
the setting using the “Q“ or “q“ soft key and confirm with “O.“<br />
y The menu for the “WP-1 INTERNAL“ device parameters is displayed.<br />
t Use the “Q“ or “q“ soft key to select and open the “Calibration/<br />
adjustment“ menu using “O.“<br />
t Use the “O“ soft key to open the “Cal/adj. sequence“ menu.<br />
t Use the “Q“ or “q“ soft key to select the “Cal. then manual<br />
adj.“ menu (factory setting) and confirm with the “l“ soft key.<br />
t Use the “o“ soft key to go to the next menu level.<br />
t Use the “O“ soft key to open the “Activate ext. adj.“ menu.<br />
t Use the “Q“ or “q“ soft key to select the “Activated“ menu item<br />
(factory setting). Not for a verifiable configuration.<br />
t Geographical data is not displayed during calibration/adjustment (factory<br />
setting).<br />
To display geographical data in the Device parameters menu, open the<br />
“Operating parameters“ submenu.<br />
Menu item “Display geogr. data > On“.