Configuring Weighing Platforms<br />
Procedure t Remove the cap.<br />
t Slide the menu access switch to the left (= “open“ position).<br />
If the device is part of a verified weighing facility, this will only be possible if<br />
the verification seal is broken. The weighing equipment must then be verified<br />
again.<br />
t Activate the Service mode, see “Service Mode.“<br />
t Select weighing platform “WP 1“ in the “Device Parameters“<br />
menu item.<br />
t If the “Internal“ setting is not already activated (marked by “o“), select<br />
the setting using the “Q“ or “q“ soft key and confirm with “O.“<br />
y The menu for the “WP-1 INTERNAL“ device parameters is displayed.<br />
t Use the “Q“ or “q“ soft key to select and open the “CaCalibration/<br />
adjustment“ menu using “O.“<br />
t Use the “Q“ or “q“ soft key to select and use “O“ to open the<br />
“Geographical data“ menu.<br />
t Use the “O“ soft key to confirm “Input parameters.“<br />
Entering Geographical Latitude and Elevation<br />
t Use the “Q“ or “q“ soft key to select the corresponding input field.<br />
t Enter the number via the keypad and confirm using the “l“ soft key.<br />
y The next input field is selected.<br />
In this example, the geographical data are entered for the respective platform as a<br />
value pair “Latitude“ and “Altitude.“ After this data was saved and the<br />
scale returned to weighing mode, this pair of values is displayed again the next time<br />
the display menu is opened. The input field for gravitational acceleration is empty.<br />
In this example, the value for gravitational acceleration is entered for the place of<br />
installation. The fields “Latitude“ and “Altitude“ are invalid. The set<br />
value is then re-displayed after it is saved and the input menu is re-opened. If you<br />
exit the Setup menu and then open the Service mode, the set value for gravity is no<br />