The displays depicted in the next two illustrations on the left show data from<br />

a multi-interval scale configured as described above, or a similarly configured<br />

multiple-range scale.<br />

If the A/D converter was configured with a “Verifiable” data record, the lines for<br />

display of metrological data (lines 1 and 2) show the data valid for use in legal<br />

metrology.<br />

The current range (e.g. R1) is displayed top left under the weighing point for<br />

multiple-range scales.<br />

ADC Configuration with Load Cell(s) Connected<br />

Configuring Weighing Platforms<br />

Procedure: 1. Open the menu access switch, see “Analog/Digital Converter (ADC)<br />

Configuration.“<br />

2. Activate the Service mode, see “Service Mode.“<br />

3. Configure WP 1, see “Analog/Digital Converter (ADC) Configuration.“<br />

4. Set single-range mode, for example, see “Analog/Digital Converter (ADC)<br />

Configuration.“<br />

5. Select the units, see “Analog/Digital Converter (ADC) Configuration.“<br />

6. Adjust without weights, see “Adjust without weights.“<br />

7. Set/Delete the preload, see “Setting the Preload“ and “Deleting the Preload.“<br />