The displays depicted in the next two illustrations on the left show data from<br />
a multi-interval scale configured as described above, or a similarly configured<br />
multiple-range scale.<br />
If the A/D converter was configured with a “Verifiable” data record, the lines for<br />
display of metrological data (lines 1 and 2) show the data valid for use in legal<br />
metrology.<br />
The current range (e.g. R1) is displayed top left under the weighing point for<br />
multiple-range scales.<br />
ADC Configuration with Load Cell(s) Connected<br />
Configuring Weighing Platforms<br />
Procedure: 1. Open the menu access switch, see “Analog/Digital Converter (ADC)<br />
Configuration.“<br />
2. Activate the Service mode, see “Service Mode.“<br />
3. Configure WP 1, see “Analog/Digital Converter (ADC) Configuration.“<br />
4. Set single-range mode, for example, see “Analog/Digital Converter (ADC)<br />
Configuration.“<br />
5. Select the units, see “Analog/Digital Converter (ADC) Configuration.“<br />
6. Adjust without weights, see “Adjust without weights.“<br />
7. Set/Delete the preload, see “Setting the Preload“ and “Deleting the Preload.“<br />
Operating Instructions <strong>Combics</strong> <strong>Complete</strong> <strong>Scale</strong>s 31