Configuring Weighing Platforms<br />
Multi-interval Mode The illustration shows an example of the input menu opened for multi-interval<br />
mode range configuration. This example shows the parameters for a scale in<br />
“Verifiable” configuration, with 2 weighing ranges and a maximum capacity of<br />
6.000 kg.<br />
– Range 1: 0…3.000 kg with e1 = 0.001 kg<br />
– Range 2: 3.002…6.000 kg with e2 = 0.002 kg<br />
t Enter the verification scale interval for Range 1 in the “E“ input field.<br />
The minimum capacity for a class l scale must be set to 0.02 kg.<br />
t Use the “o“ soft key to go to the next menu level.<br />
y The active range configuration is marked with “o.“<br />
t Use the “o“ soft key to go back to the “Verifi able“ menu.<br />
Selecting Units t Use the “q“ and “O“ soft key to open the “Available units“<br />
menu item.<br />
This menu lets you enable or disable the weight units available under “Weight<br />
unit 1,“ “Weight unit 2“ and “Weight unit 3.“<br />
t Select the respective unit using the “Q“ or “q“ soft key and confirm using the<br />
“l“ soft key.<br />
In most cases, you will not need to change defined values.<br />
Available weight units are marked by a *.<br />
The weight unit used for configuration of weighing ranges cannot be blocked.<br />
t Use the “o“ soft key to go back to the “Internal“ menu.<br />
t Use the “q“ and “O“ soft key to open the “Calibration/<br />
adjustment“ menu.<br />
t Use the “q“ and “O“ soft key to open the “Calibration/<br />
adjustment unit“ menu item to define the weight unit for<br />
calibration and adjustment. In most cases, you will not need to change defined<br />
values.<br />
y All units are displayed in the menu that are activated in “Available<br />
units.“ The current setting is marked by a “o.“<br />
Operating Instructions <strong>Combics</strong> <strong>Complete</strong> <strong>Scale</strong>s 29