Configuring Weighing Platforms<br />

Multi-interval Mode The illustration shows an example of the input menu opened for multi-interval<br />

mode range configuration. This example shows the parameters for a scale in<br />

“Verifiable” configuration, with 2 weighing ranges and a maximum capacity of<br />

6.000 kg.<br />

– Range 1: 0…3.000 kg with e1 = 0.001 kg<br />

– Range 2: 3.002…6.000 kg with e2 = 0.002 kg<br />

t Enter the verification scale interval for Range 1 in the “E“ input field.<br />

The minimum capacity for a class l scale must be set to 0.02 kg.<br />

t Use the “o“ soft key to go to the next menu level.<br />

y The active range configuration is marked with “o.“<br />

t Use the “o“ soft key to go back to the “Verifi able“ menu.<br />

Selecting Units t Use the “q“ and “O“ soft key to open the “Available units“<br />

menu item.<br />

This menu lets you enable or disable the weight units available under “Weight<br />

unit 1,“ “Weight unit 2“ and “Weight unit 3.“<br />

t Select the respective unit using the “Q“ or “q“ soft key and confirm using the<br />

“l“ soft key.<br />

In most cases, you will not need to change defined values.<br />

Available weight units are marked by a *.<br />

The weight unit used for configuration of weighing ranges cannot be blocked.<br />

t Use the “o“ soft key to go back to the “Internal“ menu.<br />

t Use the “q“ and “O“ soft key to open the “Calibration/<br />

adjustment“ menu.<br />

t Use the “q“ and “O“ soft key to open the “Calibration/<br />

adjustment unit“ menu item to define the weight unit for<br />

calibration and adjustment. In most cases, you will not need to change defined<br />

values.<br />

y All units are displayed in the menu that are activated in “Available<br />

units.“ The current setting is marked by a “o.“<br />