Analog/Digital Converter (ADC) Configuration<br />

Condition The weighing platform must already be connected.<br />

Open the Menu Access Switch The menu access switch is located on the back of the indicator right next to the<br />

weighing platform connection.<br />

t Remove the cap.<br />

t Slide the switch to the left (= “open“ position).<br />

Activating the Service Mode t See “Service Mode“ on page 17.<br />

Configuring Weighing Platforms<br />

Configuration t Select weighing platform “WP 1.“<br />

t If the “Internal“ setting is not already activated (marked by “o“), select<br />

the setting using the “Q“ or “q“ soft key and confirm with “O.“<br />

y The message “Function active“ appears briefly. The “WP 1 internal“<br />

menu will then open.<br />

t Open the ADC Confi guration menu.<br />

t Select the desired configuration data record using the “Q“ or “q“ soft key:<br />

“Standard“ or “Verifi able“ (verifiable configuration).<br />

The default setting depends on the data record.<br />

t Open the menu for configuring A/D converter parameters. In this example, the<br />

menu for ADC configuration of a weighing platform for use in legal metrology<br />

is opened.<br />

If “Standard“ was selected previously, then the “Accuracy<br />

class“ is not displayed.<br />

t Open the first menu item.<br />

For a standard configuration, the “Ranges“ menu item, for a verifiable<br />

configuration, the “Accuracy class“ menu item.<br />

When the “Verifiable” configuration is active, always select the “Accuracy<br />

class“ menu item first.<br />

t Set accuracy class l/m. The “o“ symbol should mark the setting, if required<br />

confirm using the “l“ soft key.<br />

t Press the “o“ soft key to exit the menu item.<br />

t Open the “Ranges“ menu item.<br />

Operating Instructions <strong>Combics</strong> <strong>Complete</strong> <strong>Scale</strong>s 27