Care and Maintenance<br />
h<br />
118 Operating Instructions <strong>Combics</strong> <strong>Complete</strong> <strong>Scale</strong>s<br />
Indicators are designed in compliance with European Hygienic Equipment Design<br />
Group (EHEDG) directives on suitable measures to avoid contamination, so that they<br />
are particularly easy to clean and disinfect.<br />
t Clean the indicator with a cloth lightly moistened with a soap solution.<br />
For use in the food industry, use a cleaning agent suitable for that particular<br />
working environment.<br />
t Wipe the indicator with a soft, dry cloth.<br />
Cleaning stainless steel surfaces<br />
- Only use conventional household cleaning agents which are suitable for<br />
stainless steel.<br />
- Only use solvents for cleaning stainless steel parts.<br />
t All stainless steel parts should be cleaned at regular intervals: Rub stainless steel<br />
surfaces with a moist cloth, with a cleaning agent if required, then remove all<br />
residue from the surface.<br />
t Allow device to dry. For additional protection, protective oil may be applied.<br />
Replacing the dust cover<br />
A damaged dust cover should be replaced immediately.<br />
t Remove damaged dust cover.<br />
t Place the new dust cover on the display and control unit and press it over the<br />
edge of the front and rear side of the device until it is fixed in place.<br />
Safety Inspection<br />
Safe operation of the device is no longer ensured when:<br />
– The device or the mains connecting lead shows visible damage.<br />
– The integrated power supply for the indicator no longer functions properly.<br />
– The device has been stored for a relatively long period under unfavorable<br />
conditions (e.g., excessive humidity)<br />
If there is any indication that safe operation of the device is no longer warranted:<br />
t Disconnect the power supply to the device (unplug the power cord from the<br />
mains supply) and make sure the device cannot be used for the time being.<br />
t Notify your nearest <strong>Sartorius</strong> Service Center.<br />
Maintenance and repair work may only be carried out by service technicians:<br />
– Who have access to the required maintenance documents and manuals<br />
and<br />
– Who have attended the appropriate training workshops<br />
The seals on the device indicate that the device may only be opened and maintained<br />
by authorized specialist personnel, so that the correct and safe operation of the<br />
device is ensured and the guarantee remains valid.