MAN-Combics Complete Scale CAW3P CAW3S CAH3-e - Sartorius
MAN-Combics Complete Scale CAW3P CAW3S CAH3-e - Sartorius
MAN-Combics Complete Scale CAW3P CAW3S CAH3-e - Sartorius
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
Error Codes<br />
Display Cause Solution<br />
No display elements No power present Check power supply<br />
------- Key has no function in this status<br />
Flashing k Battery defective or time changed Set time<br />
H Load exceeds weighing capacity Unload the scale<br />
L or ERR 54 Weighing pan is not in place Position the weighing pan<br />
ERR 101 - 104 Key is stuck Release key or<br />
key pressed when switching on the device Contact your local <strong>Sartorius</strong> Service Center<br />
ERR 320 Operating program memory faulty Contact your local <strong>Sartorius</strong> Service Center<br />
ERR 335 Verified weighing platform not compatible Connect a compatible weighing platform<br />
with the connected terminal<br />
ERR 340 Incorrect operating parameter (EEPROM) Turn the scale off and then on again.<br />
If the error code Err340 is still displayed<br />
, please contact your local <strong>Sartorius</strong> Service Center<br />
ERR 341 RAM has lost data Connect the device to power for at least 10 hours<br />
Battery is empty<br />
ERR 343 Loss of data in the memory area for transaction Contact your local <strong>Sartorius</strong> Service Center<br />
numbers in external alibi memory<br />
INF 01 “Display overload“ Data output not compatible with output format Set output format correctly<br />
INF 02 “Zero point error Adjustment condition was not maintained, Unload the scale first then zero, then tare via the )<br />
key at start of cal.“ e.g. not tared or weighing pan loaded<br />
INF 03 Adjustment could not be completed Allow to warm up again and repeat the adjustment<br />
within a certain time. process<br />
INF 06 “Int. weight not<br />
available or missing“<br />
Built-in adjustment weight is defective Contact your local <strong>Sartorius</strong> Service Center<br />
INF 07 “Function not allowed Function not allowed in scales verified for Contact your local <strong>Sartorius</strong> Service Center for<br />
for verifiable weighing“ use in legal metrology information on changing settings<br />
INF 08 The load on the scale is too heavy to zero Check whether “Tare/zero at power on” was complied<br />
the readout with in your configuration.<br />
INF 09 Taring is not possible when the scale gross<br />
weight is < zero<br />
Zero the scale<br />
INF 10 Tare key is blocked when there is data in the The data stored for the application program must be .<br />
tare memory deleted before taring<br />
INF 22 Error in storing reference value Load is too light, place a heavier weight on the weigher<br />
INF 23 “Application<br />
initialization error“<br />
Error in initializing an application Contact your local <strong>Sartorius</strong> Service Center<br />
INF 29 “<strong>Scale</strong> minimum load Minimum load not reached Reduce minimum load (in the “Application parameters,”<br />
not reached“ under “Minimum load for autom. initialization”)<br />
INF 71 “Invalid value,“ Cannot store the current weight value None<br />
Value too low/high“ or<br />
“selection is not possible“<br />
(e.g., control limits too low or too high)<br />
INF 72 “Maximum quantity reached“ Cannot store the current weight value<br />
(e.g. transaction counter maximum reached)<br />
None<br />
INF 73 “Memory deleted/<br />
Memory not available“<br />
Data not found or unreadable Contact your local <strong>Sartorius</strong> Service Center<br />
INF 74 “Function is not available“ Function is blocked None<br />
or<br />
“Function is blocked“<br />
(e.g. menu is blocked)<br />
INF 88 “Function was started“ A function has been activated None<br />
INF 98 No weighing platform connected Connect a weighing platform<br />
INF 99 No weighing platform connected Connect a weighing platform<br />
NO WP No weighing platform connected Connect a weighing platform<br />
116 Operating Instructions <strong>Combics</strong> <strong>Complete</strong> <strong>Scale</strong>s