MAN-Combics Complete Scale CAW3P CAW3S CAH3-e - Sartorius
MAN-Combics Complete Scale CAW3P CAW3S CAH3-e - Sartorius
MAN-Combics Complete Scale CAW3P CAW3S CAH3-e - Sartorius
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
Data Input Format<br />
Data Interfaces<br />
You can connect a computer to your scale to send commands controlling weighing<br />
instrument functions and applications via the interface port.<br />
All commands use the same data input format. They begin with the ESC (ASCII:<br />
27) character and end with a carriage return CR (ASCII: 13) and LF (ASCII: 10). The<br />
total length of a command is anywhere from 4 characters (1 command character<br />
between the start and end described above) to a max. of 7 characters (4 command<br />
characters). This number can also be higher when sending texts.<br />
The commands listed in the following table must each be supplemented with<br />
ESC ... CR LF.<br />
Example The command character for output is “P“ (“output to Port“). To trigger this<br />
command, send the string: “ESC P CR LF”.<br />
Command Meaning<br />
K Weighing mode 1<br />
L Weighing mode 2<br />
M Weighing mode 3<br />
N Weighing mode 4<br />
O Block keys<br />
P Send display value to data interface<br />
Q Output acoustic signal<br />
R Unblock keys<br />
D Tare and zero<br />
(combination tare function)<br />
f3_ Zero (see also the “kZE_” command)<br />
f4_ Tare without zeroing (see also the “kT_” command)<br />
i_ Information about the indicator, example of output: “CAI/016302/1“<br />
Meaning: Indicator: <strong>Combics</strong> 3, software version: 016302,<br />
Active weighing platform: 1<br />
kF1_ Trigger soft key F1 function<br />
kF2_ Trigger soft key F2 function<br />
kF3_ Trigger soft key F3 function<br />
kF4_ Trigger soft key F4 function<br />
kF5_ Trigger soft key F5 function<br />
kF6_ Trigger R key function<br />
kF7_ Trigger H key function<br />
kF8_ Trigger D key function<br />
kF9_ Trigger k key function<br />
kF10_ Trigger J key function<br />
kF11_ Trigger K key function<br />
kF12_ Trigger L key function<br />
kP_ Trigger p key function (print at printer interface)<br />
kT_ Trigger T key (tare)<br />
kNW_ Trigger n key function (toggle the weighing platform)<br />
kZE_ Trigger ( key function (zero the instrument)<br />
kCF_ Trigger c key function<br />
x1_ Output model designation of active weighing platform, example: “LP6200S-0C“<br />
x2_ Output serial number of active weighing platform, example: “0012345678“<br />
x3_ Output software version of active weighing platform, example: “00-43-04“<br />
x4_ Output software version of indicator, example: “01-63-02“<br />
x9_ Output serial number of indicator, example: “0012345678“<br />
x10_ Output model of indicator, example: “CAIS3“<br />
z1_ Input: printout header 1<br />
z2_ Input: printout header 2<br />
The ASCII code for the “underline” character (“_”) is 95.<br />
Format for entering printout header lines: “ESC z x a ... a _ CR LF“ with x=1 or 2<br />
and a ... a: 1 to 20 characters for header x, followed by the underline, CR and LF<br />
characters.<br />
txx…x_ Write text in display. xx…x is the text to be displayed.<br />
Operating Instructions <strong>Combics</strong> <strong>Complete</strong> <strong>Scale</strong>s 101