Data Input Format<br />

Data Interfaces<br />

You can connect a computer to your scale to send commands controlling weighing<br />

instrument functions and applications via the interface port.<br />

All commands use the same data input format. They begin with the ESC (ASCII:<br />

27) character and end with a carriage return CR (ASCII: 13) and LF (ASCII: 10). The<br />

total length of a command is anywhere from 4 characters (1 command character<br />

between the start and end described above) to a max. of 7 characters (4 command<br />

characters). This number can also be higher when sending texts.<br />

The commands listed in the following table must each be supplemented with<br />

ESC ... CR LF.<br />

Example The command character for output is “P“ (“output to Port“). To trigger this<br />

command, send the string: “ESC P CR LF”.<br />

Command Meaning<br />

K Weighing mode 1<br />

L Weighing mode 2<br />

M Weighing mode 3<br />

N Weighing mode 4<br />

O Block keys<br />

P Send display value to data interface<br />

Q Output acoustic signal<br />

R Unblock keys<br />

D Tare and zero<br />

(combination tare function)<br />

f3_ Zero (see also the “kZE_” command)<br />

f4_ Tare without zeroing (see also the “kT_” command)<br />

i_ Information about the indicator, example of output: “CAI/016302/1“<br />

Meaning: Indicator: <strong>Combics</strong> 3, software version: 016302,<br />

Active weighing platform: 1<br />

kF1_ Trigger soft key F1 function<br />

kF2_ Trigger soft key F2 function<br />

kF3_ Trigger soft key F3 function<br />

kF4_ Trigger soft key F4 function<br />

kF5_ Trigger soft key F5 function<br />

kF6_ Trigger R key function<br />

kF7_ Trigger H key function<br />

kF8_ Trigger D key function<br />

kF9_ Trigger k key function<br />

kF10_ Trigger J key function<br />

kF11_ Trigger K key function<br />

kF12_ Trigger L key function<br />

kP_ Trigger p key function (print at printer interface)<br />

kT_ Trigger T key (tare)<br />

kNW_ Trigger n key function (toggle the weighing platform)<br />

kZE_ Trigger ( key function (zero the instrument)<br />

kCF_ Trigger c key function<br />

x1_ Output model designation of active weighing platform, example: “LP6200S-0C“<br />

x2_ Output serial number of active weighing platform, example: “0012345678“<br />

x3_ Output software version of active weighing platform, example: “00-43-04“<br />

x4_ Output software version of indicator, example: “01-63-02“<br />

x9_ Output serial number of indicator, example: “0012345678“<br />

x10_ Output model of indicator, example: “CAIS3“<br />

z1_ Input: printout header 1<br />

z2_ Input: printout header 2<br />

The ASCII code for the “underline” character (“_”) is 95.<br />

Format for entering printout header lines: “ESC z x a ... a _ CR LF“ with x=1 or 2<br />

and a ... a: 1 to 20 characters for header x, followed by the underline, CR and LF<br />

characters.<br />

txx…x_ Write text in display. xx…x is the text to be displayed.<br />