Data Interfaces<br />
Setting the SBI Data Output<br />
Data output settings can be made in the Setup menu under “Data<br />
communications:SBI:Data output.“<br />
The following options are available:<br />
– The displayed value, with or without stability check<br />
– Automatic output of the displayed value, either with or without stability check,<br />
or automatically at defined intervals<br />
– Output of a printout as configured in the “Device parameters:<br />
Config. printout:Printer 1“ or “Printer 2“ (see next<br />
page).<br />
You can define the printout content by specifying which blocks of information are<br />
to be included (see “Configuring Printouts”).<br />
In general, the current display value is output (weight with unit, calculated value,<br />
alphanumeric display) unless you configure a user-definable data record.<br />
Generally, data is output only after the weighing instrument has stabilized (factory<br />
setting). Here you can define whether data is output on request or automatically,<br />
at stability or without stability check; and configure a user-defined printout.<br />
The data output for “without stability“ must be set accordingly (SBI: Data output).<br />
If you select time-dependent automatic printout, you need to define the print<br />
interval (in display updates) as well.<br />
Each line of a printout can contain up to 20 characters. The first 6 characters,<br />
called the “data header”, identify the subsequent value. Only 14 characters are<br />
available if you disable the header. This is carried out in the “Line format“ menu<br />
item (see also “Setup Overview“ in “Configuration“).<br />
Automatic Data Output (SBI) You can have the weight readout printed automatically 1 ). This printout can be<br />
generated after a certain number of display updates 2 ). You can also configure<br />
whether or not the auto-print function is dependent on the stability of the scale 3 ).<br />
The display update frequency depends on both the scale model and the operating<br />
status.<br />
Examples:<br />
N + 153.00 g Net weight<br />
Stat Display blank<br />
Stat L Display underload<br />
Stat H Display overload<br />
100 Operating Instructions <strong>Combics</strong> <strong>Complete</strong> <strong>Scale</strong>s<br />
„Data output“ setting:<br />
1) 3) Automatic, without stability or automatic with stability.<br />
Factory setting: Manual after stability; i.e., automatic data output function off.<br />
2) Time-dependent automatic data output:<br />
Interval: 1, 2, 10 or 100 display updates<br />
Factory setting: 1 display update