Data Interfaces<br />

Setting the SBI Data Output<br />

Data output settings can be made in the Setup menu under “Data<br />

communications:SBI:Data output.“<br />

The following options are available:<br />

– The displayed value, with or without stability check<br />

– Automatic output of the displayed value, either with or without stability check,<br />

or automatically at defined intervals<br />

– Output of a printout as configured in the “Device parameters:<br />

Config. printout:Printer 1“ or “Printer 2“ (see next<br />

page).<br />

You can define the printout content by specifying which blocks of information are<br />

to be included (see “Configuring Printouts”).<br />

In general, the current display value is output (weight with unit, calculated value,<br />

alphanumeric display) unless you configure a user-definable data record.<br />

Generally, data is output only after the weighing instrument has stabilized (factory<br />

setting). Here you can define whether data is output on request or automatically,<br />

at stability or without stability check; and configure a user-defined printout.<br />

The data output for “without stability“ must be set accordingly (SBI: Data output).<br />

If you select time-dependent automatic printout, you need to define the print<br />

interval (in display updates) as well.<br />

Each line of a printout can contain up to 20 characters. The first 6 characters,<br />

called the “data header”, identify the subsequent value. Only 14 characters are<br />

available if you disable the header. This is carried out in the “Line format“ menu<br />

item (see also “Setup Overview“ in “Configuration“).<br />

Automatic Data Output (SBI) You can have the weight readout printed automatically 1 ). This printout can be<br />

generated after a certain number of display updates 2 ). You can also configure<br />

whether or not the auto-print function is dependent on the stability of the scale 3 ).<br />

The display update frequency depends on both the scale model and the operating<br />

status.<br />

Examples:<br />

N + 153.00 g Net weight<br />

Stat Display blank<br />

Stat L Display underload<br />

Stat H Display overload<br />

100 Operating Instructions <strong>Combics</strong> <strong>Complete</strong> <strong>Scale</strong>s<br />

„Data output“ setting:<br />

1) 3) Automatic, without stability or automatic with stability.<br />

Factory setting: Manual after stability; i.e., automatic data output function off.<br />

2) Time-dependent automatic data output:<br />

Interval: 1, 2, 10 or 100 display updates<br />

Factory setting: 1 display update