Maintenance<br />

3-16<br />

• Complete is flashing<br />

• There are three keys to choose from EXIT, REPEAT, and FACTOR.<br />

EXIT – If the operator chooses EXIT, the system acknowledges the<br />

Material Factor is not used. The system does not ask for a material<br />

factor any more for this calibration method until a manual span entry is<br />

performed. By pressing EXIT, the operator tells the system that<br />

material factor is not desired, but wants to use the test results for<br />

changing the span number. Proceed to Section 3.10.3.1.6.<br />

REPEAT – If the operator chooses REPEAT, proceed to Section<br />

3.10.3.1.4.<br />

FACTOR – If the operator chooses FACTOR, the following screen<br />

displays:<br />

XXX Matl FACTOR<br />

New factor: 000.00 %<br />

Change factor?<br />

YES NO ADV<br />

ADV – advances to Old Factor and again to New Factor.<br />

NO – the Material Factor is set to 1.00<br />

YES – the following screen displays<br />

XXX Matl FACTOR<br />

Old factor: 000.00 %<br />

New factor: 000.00 %<br />

RUN MENU REPEAT<br />

REPEAT – returns the operator to Section 3.10.3.1.4<br />

RUN – returns to the RUN menu main screen<br />

MENU – returns to MAIN MENU 1<br />

XXX = R-Cal, WTS, or CHAINS<br />

XXX = R-Cal, WTS, or CHAINS<br />

Maintenance <strong>Thermo</strong> Fisher Scientific<br />

REC 4175 Rev B