Maintenance<br />

3-14<br />

3.10.3.1.3 Starting a Test Weights Calibration<br />

Use the following steps to begin a Test Weights Calibration:<br />

1. Stop the conveyor belt if necessary to apply test weights.<br />

2. Press SPAN CAL<br />

The following screen displays:<br />

AUTO SPAN Weights<br />

Press START to begin<br />

weight calibration.<br />

START EXIT MANUAL<br />

3. Press START to disengage the Master Tons counter.<br />

The following screen displays:<br />

AUTO SPAN Weights<br />

Apply weights, then<br />

press START.<br />

START EXIT MANUAL<br />

If LOAD WTS output was selected in the I/O Definition Scroll, test weights<br />

are automatically loaded after START is pressed. Wait for test weights to be<br />

loaded before proceeding. The following message displays:<br />

AUTO SPAN Weights<br />

Press START to load<br />

test weights.<br />

START EXIT<br />

EXIT returns the screen to MAIN MENU 1<br />

4. Apply test weights<br />

5. Press START<br />

The following screen displays:<br />

AUTO SPAN Weights<br />

Run belt, then<br />

press START.<br />

START EXIT<br />

Password: Operator<br />

Password: Operator<br />

Password: Operator<br />

Maintenance <strong>Thermo</strong> Fisher Scientific<br />

REC 4175 Rev B