<strong>Thermo</strong> Fisher Scientific Operation<br />

REC 4175 Rev B<br />

Operation<br />

The problem causing the calibration to be aborted must be corrected before<br />

attempting another calibration. The information entered previously is still in<br />

memory and should not have to be reentered. After the problem has been<br />

corrected, press MENU. Main Menu 1 should appear. If not, continue to press<br />

MENU until Main Menu 1 appears.<br />

Proceed to Section 3.9, MAIN MENU 1, and conduct a Zero Calibration from<br />

3.10.1 and a Span Calibration from Section 3.10.3. The scale is now calibrated<br />

using R-Cal as the simulated load. Press RUN to start or MENU for scrolls<br />

2-45