Operation<br />

2-42<br />

When duration test is finished the new values for length of belt and time are<br />

displayed. Press CONITNUE.<br />

TEST DURATION<br />

Length = 0000.0 Ft<br />

Time = 000 sec<br />

START ABORT<br />

During the Acquire Test Duration, the Integrator performed an unassisted zero<br />

calibration and installed the new zero. Skip to Step 19.<br />

2.9.8 Manual Entry of Test Duration<br />

This procedure allows direct entry of parameters that would otherwise be<br />

generated by the acquire Test Duration modes. This menu is generally used<br />

when the operator cannot see the belt while standing at the front panel.<br />

If MANUAL is pressed, the system prompts the operator for running the belt at<br />

its maximum speed. Then press CONTINUE<br />

Pressing ABORT returns the screen to Cal Data Scroll 11.<br />

Start Belt. Press<br />

CONTINUE when belt<br />

is at maximum speed.<br />

ABORT CONTINUE<br />

The operator is prompted to enter the length of one belt revolution.<br />

Pressing ABORT returns the screen to Cal Data Scroll 11.<br />

ENTER length of one<br />

Belt revolution<br />

Length 000.0 Ft<br />

ENTER ABORT<br />

If English Default: 200.0 Ft<br />

or Mixed: Min: 1.0 Ft<br />

Max: 10000.0 Ft<br />

If Metric Default: 200.0 m<br />

Min: 1.0 m<br />

Max: 1000.0 m<br />

The number of belt revolutions to be timed is than entered.<br />

Pressing ABORT returns the screen to Cal Data Scroll 11.<br />

Operation <strong>Thermo</strong> Fisher Scientific<br />

REC 4175 Rev B