<strong>Thermo</strong> Fisher Scientific Operation<br />

REC 4175 Rev B<br />

Operation<br />

2.9.7 Partial Test Duration Acquisition<br />

If PARTIAL is pressed, the operator is asked to enter the length between two<br />

marks on the belt.<br />

Pressing ABORT returns the screen to Cal Data Scroll 11.<br />

Enter length between<br />

two marks on belt.<br />

Length: 200.0 Ft<br />

ENTER ABORT<br />

If English Default: 200 Ft<br />

or Mixed: Min: 1.0 Ft<br />

Max: 10000.0 Ft<br />

If Metric: Default: 50.0 m<br />

Min: 0.5 m<br />

Max: 3000.0 m<br />

After the length of the belt has been entered, the system automatically moves to<br />

the following screen, which prompts the operator to press START when the<br />

mark passes the reference point.<br />

The belt must be running at the maximum speed before executing this function.<br />

If not, the prescaler is improperly calculated.<br />

Pressing ABORT returns the screen to Cal Data Scroll 11.<br />

Start belt. Press<br />

START when 1 st mark<br />

passes reference<br />

START ABORT<br />

When START is pressed above, the system automatically moves to the next<br />

instructional screen. The operator presses COUNT each time a mark passes the<br />

reference point until both marks have passed. Then press DONE.<br />

Pressing ABORT returns the screen to Cal Data Scroll 11.<br />

Zero Cal light illuminates indicating zero calibration in progress.<br />

Press COUNT each<br />

Time a mark passes.<br />

00000 secs 000 revs<br />

START ABORT<br />

2-41