<strong>Thermo</strong> Fisher Scientific Operation<br />

REC 4175 Rev B<br />

Operation<br />

2.6.1.3.4 Pivot to 2nd Weigh Idler (Optional)<br />

As indicated in Figure 2-11, measure the distance from the pivot centerline to<br />

the second weigh idler’s centerline to the nearest 0.032 (1/32”). Record the<br />

distance below.<br />

(Pivot to 2 nd Weigh Idler Distance)<br />

(Model 10-22 belt scale, 10-17-2 similar.)<br />

Figure 2-11: Pivot to 2nd Weigh Idler<br />

Note: Up to six (6) weigh idlers on a single weighbridge may be entered during<br />

programming. Measure and record the same as Section 2.6.1.3.4 for each<br />

additional weigh idler.<br />

2.6.1.3.5 Pivot to 1 st , 2 nd , 3 rd and 4 th Weigh Idler<br />

As indicated in Figure 2-12, measure the distance from the pivot centerline to<br />

the weigh idler’s centerline to the nearest 0.032 (1/32”). Record the distance<br />

below.<br />

(Pivot to Weigh Idler Distance)<br />

Figure 2-12: Pivot to 1st, 2nd, 3rd, and 4th Weigh Idler<br />

2.6.1.4 Calibration Using Test Weights<br />

Complete the steps in Sections 2.6.1.4.1, 2.6.1.4.2, 2.6.1.4.3, and 2.6.1.4.4<br />

below only if test weights are used for calibration.<br />

2-19