Operation<br />

2-8<br />

2.4.2 Connecting Incoming Power – Panel Mount<br />

To connect incoming power for panel mount installation, use the following<br />

procedure.<br />

Note: All units shipped from the factory are configured for 115 VAC. If 230<br />

VAC is desired, refer to Section 2.5.1, motherboard configuration Jumpers and<br />

Switches.<br />

• For input power, use 14 AWG standard wire<br />

• Wire the safety ground terminal located on the right backside of the<br />

enclosure.<br />

• Wire the HOT to terminal labeled 1 of Terminal L on the Power Input<br />

Terminal.<br />

• Wire the NEUTRAL to the terminal labeled 2 of Terminal N on the Power<br />

Input Terminal.<br />

2.5 Integrator Configuration<br />

The Micro-Tech 3101 is one of a family of products that is supported by a<br />

common hardware platform. Configuration of the hardware platform and<br />

additional circuit boards enable the shareware platform to be used for several<br />

discrete instruments.<br />

Wire jumpers are installed at the factory for the instrument ordered and should<br />

not have to be reconfigured in the field.<br />

Switches and removable jumpers are described in this section. The default<br />

position is noted in each description and, in most cases, is not changed.<br />

2.5.1 Mother Board Configuration Jumpers and Switches<br />

TO BE PERFORMED BY QUALIFIED SERVICE PERSONNEL ONLY.<br />

1. General Purpose Digital Inputs<br />

Located on the motherboard are provisions for 5 programmable status<br />

inputs. The programmable inputs may be configured as normally open or<br />

normally closed. Inputs are desigend for dry contacts.<br />

Input 1 may also be used as a speed senor input. Refer to Table 2-1 for<br />

configuration inofrmation.<br />