<strong>Thermo</strong> Fisher Scientific Operation<br />

REC 4175 Rev B<br />

Operation<br />

Figure 2-2: Electrical and Mounting Guidelines of the Micro-Tech Model 3101 (Field Mount)<br />

Integrator<br />

CAUTION<br />

REFER TO THE FILED WIRING DIAGRAM AS A GUIDE IF YOU<br />

DO NOT HAVE A SPECIFIC WIRING DIAGRAM FOR YOUR<br />

SYSTEM. FOLLOW YOUR LOCAL ELECTRICAL CODES AND<br />

REGUATIONS FOR MINIMUM WIRE SIZE AND ROUTING.<br />

2.3.2 Connecting Incoming Power - Field Mount<br />

To connect the incoming power, use the following procedure.<br />

Note: All units shipped from the factory are configured for 115 VAC. If you<br />

desired 230 VAC make sure the power selector switch is set to 230 VAC<br />

(Section 2.5.1).<br />

1. Loosen the screw latch mounted on the front chassis. Open the door (see<br />

Figure 2-3).<br />

2. Route incoming power wiring through a conduit hole at the bottom right of<br />

the enclosure. Leave ample loose wiring (typically 8”) to facilitate<br />

removing the terminal connectors.<br />

3. Wire safety ground terminal located on the side of the chassis.<br />

4. Wire HOT to Terminal L on Power Input Terminal.<br />

5. Wire NEUTRAL to Terminal N on Power Input Terminal.<br />

6. If additional I/O is required at the line voltages, these wires should be routed<br />

through a conduit hole on the bottom right of the enclosure. Leave ample<br />

loose wiring (typically 8”) to facilitate removing the terminal connectors.<br />

2-5