Minor Calibration **** THIS PAGE LEFT BLANK INTENTIONALLY **** E-4 Minor Calibration Thermo Fisher Scientific REC 4175 Rev B
Appendix F Minor Calibration Notes: 1. Run belt empty during Zero & Span Calibration 2. Always perform a Zero Calibration prior to a Span Calibration 3. Record Percent Error End at the end of Step 3.0 4. Record Old and new Zero and Span numbers is Step 4 = Zero Calibration = = Span Calibration = Step Function Display Step Function Display 1.0 Press MENU 2.0 Press ZERO CAL MAIN MENU 1 ZERO CAL 2.0 AUTO 3.0 Press START ZEROING =Time Remaining Counts Down to Zero = When Zero Upper Display Auto Zero Complete “Complete” Flashing 1.0 Press MENU Pres SPAN CAL 3.0 Press START Thermo Fisher Scientifc Optional Documentation REC 4175 Rev B MAIN MENU 1 AUTO SPAN R CAL AUTO SPANNING =Time Remaining Counts Down When Zero Upper Display Auto Span Complete “Complete” Flashing 4.0 Press YES 4.0 Press YES 5.0 Press RUN 5.0 Press RUN Zero Calibration Complete Span Calibration Complete To abort during Zero or Span Calibration, press RUN. F-1
- Page 193 and 194: Menus WARNING CHANGING THE DEFINITI
- Page 195 and 196: Menus - I/O DEF SCROLL 9A- BCD inpu
- Page 197 and 198: Menus A.11.1. Define Rate Alarm Use
- Page 199 and 200: Menus - ALARM SCROLL 3A - Low speed
- Page 201 and 202: Menus - COMM. A SCROLL 4 - Wordleng
- Page 203 and 204: Menus Default: 8 BITS Selections: 7
- Page 205 and 206: Menus A form feed character can be
- Page 207 and 208: Menus There are three ways for defi
- Page 209 and 210: Menus Once you have defined the str
- Page 211 and 212: Menus The Reset Total: - PRINTER SC
- Page 213 and 214: Menus A.12.5. The PRINT Key The PRI
- Page 215 and 216: Menus A.13. MAIN MENU 6 Main Menu 6
- Page 217 and 218: Appendix B Weighbridge Physical Par
- Page 219 and 220: Digital Input/Output Appendix Figur
- Page 221 and 222: Digital Input/Output # MODEL BELT W
- Page 223 and 224: Digital Input/Output # MODEL BELT W
- Page 225 and 226: Appendix C Linearization Request RE
- Page 227 and 228: Appendix D Digital Input/Output The
- Page 229 and 230: Digital Input/Output D.2. Digital I
- Page 231 and 232: Digital Input/Output Thermo Fisher
- Page 233 and 234: Digital Input/Output Appendix Figur
- Page 235 and 236: Digital Input/Output D.4. Analog I/
- Page 237 and 238: Digital Input/Output Appendix Figur
- Page 239 and 240: Digital Input/Output Appendix Figur
- Page 241 and 242: Appendix E Multi-Point Calibration
- Page 243: Multi-Point Calibration E.2.2. Digi
- Page 247 and 248: Appendix G Optional Documentation T
- Page 249 and 250: Appendix H Engineering Drawings Fi
Appendix F<br />
Minor Calibration<br />
Notes:<br />
1. Run belt empty during Zero & Span Calibration<br />
2. Always perform a Zero Calibration prior to a Span Calibration<br />
3. Record Percent Error End at the end of Step 3.0<br />
4. Record Old and new Zero and Span numbers is Step 4<br />
= Zero Calibration = = Span Calibration =<br />
Step Function Display Step Function Display<br />
1.0 Press MENU<br />
2.0<br />
Press ZERO<br />
CAL<br />
MAIN<br />
MENU 1<br />
ZERO CAL 2.0<br />
AUTO<br />
3.0 Press START<br />
ZEROING<br />
=Time Remaining Counts Down to Zero =<br />
When Zero<br />
Upper Display Auto Zero Complete<br />
“Complete” Flashing<br />
1.0 Press MENU<br />
Pres SPAN<br />
CAL<br />
3.0 Press START<br />
<strong>Thermo</strong> Fisher Scientifc Optional Documentation<br />
REC 4175 Rev B<br />
MAIN<br />
MENU 1<br />
AUTO<br />
SPAN R<br />
CAL<br />
AUTO<br />
SPANNING<br />
=Time Remaining Counts Down<br />
When Zero<br />
Upper Display Auto Span Complete<br />
“Complete” Flashing<br />
4.0 Press YES 4.0 Press YES<br />
5.0 Press RUN 5.0 Press RUN<br />
Zero Calibration Complete Span Calibration Complete<br />
To abort during Zero or Span Calibration, press RUN.<br />
F-1