Multi-Point Calibration<br />

E.2.2. Digital Input Selection<br />

When the number of calibrations selected is greater than 1, scrolls are added to<br />

the Digital Input Define (DIG. INPUT DEF) scroll in I/O DEF SCROLL 4 as<br />

indicated below:<br />

-- I/O DEF SCROLL 4 --.<br />

Dig. Input Def.<br />

CAL. Sel 2 0 NC<br />

ENTER NC/NO NEXT<br />

The number of Cal Sel inputs is equal to the number of calibration<br />

numbers. When an input is activated, the corresponding calibration number and<br />

associated zero, span, and Linearization table is also activated. In the event<br />

more than one input is activated, the first input active is used.<br />

When an auto zero or auto span is executed, the results are saved to the active<br />

calibration number. Once a calibration test has been initiated, changing the<br />

active calibration number via the digital inputs has no effect; the calibration<br />

factors, zero, or span is saved to the calibration number that was active at the<br />

beginning of the test.<br />

E.2.3. Linearization<br />

In the Linearization scroll, the table of Linearization factors displayed is only for<br />

the active calibration number.<br />

<strong>Thermo</strong> Fisher Scientific Multi-Point Calibration E-3<br />

REC 4175 Rev B