Appendix E<br />

Multi-Point Calibration<br />

This appendix contains setup and calibration menus for establishing multiple<br />

calibration factors, both zero and span, for the weigh system.<br />

Variations in conveyor belt tensions affect the accuracy capabilities of a<br />

conveyor belt scale. Causes of these variations may be the use of a reversing<br />

belt, multiple feed points, or multiple products being conveyed that have a<br />

significant difference in belt density.<br />

To compensate for these changes in the system operation, and to maintain<br />

weighing accuracy under varying mechanical influences, different calibration<br />

factors may be used.<br />

Up to 10 different calibration points are possible. Each calibration point has its<br />

own zero number, span number, and Linearization table. To configure the<br />

Linearization table, refer to REC 3909. The different calibration points are<br />

accessible through the front panel keyboard or via a digital input. When a<br />

calibration number is selected, all of its associated values immediately become<br />

active.<br />

E.1. Define Number of Calibration Points<br />

During the initial programming (cold start) of the Integrator, a theoretical zero<br />

and span are computed and applied to all of the calibration points. From this<br />

point on, any zero or span calibration performed changes only the currently<br />

active calibration point.<br />

Enabling multiple calibration points is performed at the CAL DATA SCROLL.<br />

Use the following steps to enable multiple calibration points.<br />

1. Select MENU<br />

2. Select MENU<br />

3. Select CALIB<br />

4. Select DATA<br />

5. Scroll down until the following screen displays:<br />

Default: 1<br />

Min: 1<br />

Max: 10<br />

-- CAL DATA SCROLL 14 --<br />

Number of calib.<br />

1<br />

ENTER<br />

Password:<br />

Operator<br />

<strong>Thermo</strong> Fisher Scientifc Minor Calibration<br />

REC 4175 Rev B<br />

E-1