Menus<br />

A.13. MAIN MENU 6<br />

Main Menu 6 is dedicated to Audit Trails and Linearization.<br />

A.13.1. Audit Trail<br />

This menu is only displayed if the Audit Trails option is installed.<br />

Audit trail meets NIST HB 44 Category 3: Remote configuration capability,<br />

access may be unlimited or controlled through a password. It consists of an<br />

event logger that includes an event counter (000 to 999), the parameter<br />

description, the date, and time of the change, and the old and new value of the<br />

parameter. Parameters may be configuration parameters or routine calibration.<br />

Events and changes may be viewed on the Integrator's display or printed out by<br />

an on-site printer.<br />

The Logger records all parameters affecting the calibration of the scale. It also<br />

records when calibration was performed. The Event Counter increments one<br />

count for each event.<br />

Audit trail records the time and displays the new and old data for any change in<br />

the items below, indicating each by an event number:<br />

• ZERO<br />

• SPAN<br />

• CALIBRATION METHOD<br />

• CHAIN WEIGHT<br />

• TEST WEIGHT<br />

• SCALE CAPACITY<br />

• SCALE DIVISION<br />

• SCALE MODEL NO.<br />

• IDLER SPACING<br />

• INCLINATION<br />

• LOAD CELL SIZE<br />

• LOAD CELL SENSITIVITY<br />

• LOAD CELL RESISTANCE<br />

• SPEED INPUT TYPE<br />

• DEAD BAND<br />

• MEASURE UNITS<br />

• REMOTE COUNTER DIVIDER<br />

• REMOTE COUNTER DURATION<br />

• AZT TRACKING RANGE<br />

NOTE: Changes in zero by auto zero tracking are not recorded. However, if<br />

AZT limit is reached, is recorded.<br />

• LINEARIZATION<br />

<strong>Thermo</strong> Fisher Scientific Weighbridge Physical Parameters A-81<br />

REC 4175 Rev B