Menus<br />

- ALARM SCROLL 3A -<br />

Low speed set<br />

10 % 10 sec<br />

ENTER SET/DELAY UNITS/%<br />

Default: 10% 10 sec<br />

Min: 0 % 0 sec<br />

Max: 105 % 90 sec<br />

- ALARM SCROLL 3B -<br />

High speed set<br />

100 % 10 sec<br />

ENTER SET/DELAY UNITS/%<br />

Default: 100% 10 sec<br />

Min: 0 % 0 sec<br />

Max: 150 % 90 sec<br />

A.11.3.1 Setup Alarm Modes<br />

The following message is displayed for three seconds.<br />

- ALARM SCROLL 5<br />

- ALARM DEFINITION -<br />

Use NEXT key or<br />

enter alarm number<br />

Password: Operator<br />

Password: Operator<br />

After three seconds, the ALARM screen is displayed. The user can use the<br />

CHOICE soft key to select the desired mode between ALARM (just a warning<br />

message), SHUT DOWN (Warning plus fault output) and NONE (no action).<br />

Confirm with ENTER. Use the NEXT key to scroll between alarms, or enter the<br />

alarm number.<br />

Alarms are listed in Section 4.3.<br />

<strong>Thermo</strong> Fisher Scientific Weighbridge Physical Parameters A-65<br />

REC 4175 Rev B