Menus<br />

WARNING<br />

CHANGING THE DEFINITION OF THE DIGITAL<br />

OUTPUTS MAY CAUSE MACHINERY TO START AFTER<br />

THE USER TRIES TO CHANGE A DEFINITION. THE<br />

FOLLOWING MESSAGE IS DISPLAYED.<br />

IF THE USER PRESSES CONTINUE, BE AWARE THE<br />

ACTION MAY CAUSE DAMAGE OR INJURY. IF THE<br />

USER PRESSES ABORT, THE SYSTEM RETURNS TO<br />

THE PREVIOUS SCROLL.<br />

A.10.7. Define Remote Totalizer Output<br />

If the TOTALIZER output function is assigned to an output, the following<br />

screens allow the user to set up related parameters. Set the divider according to<br />

the maximum rate the scale will run. The divider is entered in Totalization units<br />

(T.U.). The pulse frequency generated in normal conditions should not exceed<br />

10 Hz. Higher frequencies are possible, however they do not improve accuracy.<br />

Default: 1<br />

Min: 0.01<br />

Max: 100<br />

Enter the pulse width in seconds of the Totalizer. A higher pulse width limits<br />

the maximum frequency. The default 0.1 sec is recommended for frequencies<br />

lower than 5 Hz.<br />

Default: 0.1 sec<br />

Min: 0.005 sec<br />

Max: 1 sec<br />

WARNING<br />

EQUIPMENT MAY START<br />

CONTINUE ABORT<br />

- I/O DEF SCROLL 6 -<br />

Remote counter div.<br />

1<br />

ENTER<br />

- I/O DEF SCROLL 7 -<br />

Remote counter pulse<br />

width 0.100 sec<br />

ENTER<br />

Password: Service<br />

Password: Service<br />

<strong>Thermo</strong> Fisher Scientific Weighbridge Physical Parameters A-59<br />

REC 4175 Rev B