Menus Default: OFF Selections: OFF, INCLIN (ATION), MOISTURE Default: OFF Selections: OFF, INCLIN(ATION), MOISTURE A.10.3. Setup Moisture Compensation Input If an analog input has been programmed for reading the moisture signal and CALIB was pressed, the following screens appear. The user can calibrate the input signal by entering the equivalence between percent of moisture and voltage on two points. Use the %Moist key to enter the percent of moisture, use the volt key to enter the corresponding number of volts, and confirm with ENTER. Default: 0.0 % 0.0 V Min: 0.0 % 2.0 V Max: 20.0 % 2.5 V Do the same with the second point shown below. Default: 12.0 % 5.0 V Min: 1.0 % 1.0 V Max: 20.0 % 5.0 V - I/O DEF SCROLL 3 - Analog Input #2 def. > Moisture < CHOICES ENTER CALIB - I/O DEF SCROLL 2A- Moisture input calibr. 0.0 %M = 2.0 V ENTER %Moist Volt - I/O DEF SCROLL 2B- Moisture input calibr. 20.0 %M = 5.0 V ENTER %Moist Volt Password: Service Password: Service Password: Service A-52 Weighbridge Physical Parameters Thermo Fisher Scientific REC 4175 Rev B
Menus A.10.4. Setup Inclination Compensation Input If an analog input has been programmed for reading the angle signal and CALIB was pressed, the following screens appear. The user can calibrate the input signal by entering the number of degrees corresponding to the minimum signal (0 mA or 4 mA) and the number of degrees corresponding to the maximum signal (20 mA). Use the +/- key to enter negative numbers, and confirm with ENTER. - I/O DEF SCROLL 3A- Conveyor low posit 00.0 Degrees ENTER +/- Default: 0.0 Degrees Min: - 25 Degrees Max: + 25 Degrees - I/O DEF SCROLL 3B- Conveyor high posit 00.0 Degrees ENTER +/- Default: 0.0 Degrees Min: - 25 Degrees Max: + 25 Degrees Note: Incline angles greater than 16° are not recommended and generally restrict the system accuracy. Consult the factory on angles with installation greater than 16°. A.10.5. Define Digital Inputs Digital inputs can be programmed. The following screen shows one logical function per time, and allows the user to assign it to a physical input. The NEXT key scrolls between the logical functions. The NC/NO key selects the Normally Open (NO) or Normally Closed (NC) status of the input. Normally Open means the input is inactive when disconnected. To program a function, scroll with NEXT until the function is displayed, then enter the physical input number and confirm with ENTER; finally scroll with NC/NO until the desired mode is displayed. By assigning a function to 0, the function is disabled. - I/O DEF SCROLL 4 - Dig. Input def. Ext Alarm 1 0 NC ENTER NC/NO NEXT Password: Service Password: Service Password: Service The following table shows the available logical selections that can be assigned to any available physical input. Typical field wiring drawings and customer specific field wiring drawings show Belt Running (Conveyor Run) defaulted to Thermo Fisher Scientific Weighbridge Physical Parameters A-53 REC 4175 Rev B
- Page 135 and 136: A.1. Menu Displays Appendix A Menus
- Page 137 and 138: Menus EXIT: returns to Main Menu 1
- Page 139 and 140: Menus When ADV is pressed, the syst
- Page 141 and 142: Menus A.3.3.3 Starting a Test Weigh
- Page 143 and 144: Menus 1. Complete is flashing 2. Th
- Page 145 and 146: Menus A.3.3.7 Ending an Auto Span P
- Page 147 and 148: Menus 4. Press CONTINUE The followi
- Page 149 and 150: Menus If R-CAL, WEIGHTS, or CHAINS
- Page 151 and 152: Menus A.5.3. Length Units -- DISPLA
- Page 153 and 154: Menus A.5.9. Displayed Damping Fact
- Page 155 and 156: Menus -- SC DATA SCROLL 3 -- Belt s
- Page 157 and 158: Menus Only if # of weigh idlers ≥
- Page 159 and 160: Menus Carry Roll to Carriage Height
- Page 161 and 162: Menus A.6.6.3 Load Cell Resistance
- Page 163 and 164: Menus A.7.1.1 Defining the R-Cal Pa
- Page 165 and 166: Menus The scroll displays the date
- Page 167 and 168: Menus When START is pressed, the sy
- Page 169 and 170: Menus When test duration is finishe
- Page 171 and 172: Menus The following scrolls are onl
- Page 173 and 174: Menus A SERVICE password is require
- Page 175 and 176: Menus Default: 3497 Min: 0 Max: 300
- Page 177 and 178: Menus A.9.5. Display Software Versi
- Page 179 and 180: Menus A.9.8. Force Cold Start This
- Page 181 and 182: Menus WARNING IF THE USER PRESSES C
- Page 183 and 184: Menus A.9.18. Test BCD Output Board
- Page 185: Menus Default: (1st) 0 sec, (2nd) 0
- Page 189 and 190: Menus Appendix Table A-4: Mother Bo
- Page 191 and 192: Menus Appendix Table A-6: Available
- Page 193 and 194: Menus WARNING CHANGING THE DEFINITI
- Page 195 and 196: Menus - I/O DEF SCROLL 9A- BCD inpu
- Page 197 and 198: Menus A.11.1. Define Rate Alarm Use
- Page 199 and 200: Menus - ALARM SCROLL 3A - Low speed
- Page 201 and 202: Menus - COMM. A SCROLL 4 - Wordleng
- Page 203 and 204: Menus Default: 8 BITS Selections: 7
- Page 205 and 206: Menus A form feed character can be
- Page 207 and 208: Menus There are three ways for defi
- Page 209 and 210: Menus Once you have defined the str
- Page 211 and 212: Menus The Reset Total: - PRINTER SC
- Page 213 and 214: Menus A.12.5. The PRINT Key The PRI
- Page 215 and 216: Menus A.13. MAIN MENU 6 Main Menu 6
- Page 217 and 218: Appendix B Weighbridge Physical Par
- Page 219 and 220: Digital Input/Output Appendix Figur
- Page 221 and 222: Digital Input/Output # MODEL BELT W
- Page 223 and 224: Digital Input/Output # MODEL BELT W
- Page 225 and 226: Appendix C Linearization Request RE
- Page 227 and 228: Appendix D Digital Input/Output The
- Page 229 and 230: Digital Input/Output D.2. Digital I
- Page 231 and 232: Digital Input/Output Thermo Fisher
- Page 233 and 234: Digital Input/Output Appendix Figur
- Page 235 and 236: Digital Input/Output D.4. Analog I/
Menus<br />
Default: OFF<br />
Selections: OFF, INCLIN (ATION), MOISTURE<br />
Default: OFF<br />
Selections: OFF, INCLIN(ATION), MOISTURE<br />
A.10.3. Setup Moisture Compensation Input<br />
If an analog input has been programmed for reading the moisture signal and<br />
CALIB was pressed, the following screens appear. The user can calibrate the<br />
input signal by entering the equivalence between percent of moisture and<br />
voltage on two points. Use the %Moist key to enter the percent of moisture,<br />
use the volt key to enter the corresponding number of volts, and confirm with<br />
ENTER.<br />
Default: 0.0 % 0.0 V<br />
Min: 0.0 % 2.0 V<br />
Max: 20.0 % 2.5 V<br />
Do the same with the second point shown below.<br />
Default: 12.0 % 5.0 V<br />
Min: 1.0 % 1.0 V<br />
Max: 20.0 % 5.0 V<br />
- I/O DEF SCROLL 3 -<br />
Analog Input #2 def.<br />
> Moisture <<br />
CHOICES ENTER CALIB<br />
- I/O DEF SCROLL 2A-<br />
Moisture input calibr.<br />
0.0 %M = 2.0 V<br />
ENTER %Moist Volt<br />
- I/O DEF SCROLL 2B-<br />
Moisture input calibr.<br />
20.0 %M = 5.0 V<br />
ENTER %Moist Volt<br />
Password: Service<br />
Password: Service<br />
Password: Service<br />