Menus<br />

A.9.15. Test Communication A<br />

The following screen allows checking the installed serial lines using a loop back<br />

type test. The maximum line number is 2; Transmit must be tied to receive for<br />

this test. "Port 1" is standard and "Port 2" is shown only if optional Comm<br />

boards are detected.<br />

- TEST SCROLL 8 -<br />

Test Communication A<br />

Port 1 Port 2<br />

By pressing the PORT 1 or the PORT 2 soft key, the test is initiated. A test<br />

pattern is sent out on the TX output and read on the RX input. If the test fails,<br />

the message "Test Failed" is shown; otherwise, the message "Test Passed" is<br />

displayed.<br />

A.9.15.1 Test RS232<br />

To test RS232 the test requires a hardware jumper to be installed between<br />

terminals TB3-22 (RX) and TB3-21 (TX).<br />

A.9.15.2 Test RS485<br />

To test the RS485 the test requires a hardware jumper to be installed between<br />

terminals TB3-21 and TB3-28 to TB3-20 and TB3-29.<br />

A.9.16. Test Communication B<br />

This test is similar to the previous one but works for the field bus version of the<br />

communication board.<br />

- TEST SCROLL 9 -<br />

Test communication B<br />

START<br />

A.9.17. Test BCD Input Board<br />

If an optional BCD input card is detected, the following screen appears. The test<br />

is similar to the digital input test.<br />

- TEST SCROLL 10 -<br />

BCD Input test<br />

0000<br />

Password: Service<br />

A-48 Weighbridge Physical Parameters <strong>Thermo</strong> Fisher Scientific<br />

REC 4175 Rev B