WARNING<br />

IF THE USER PRESSES CONTINUE, BE AWARE<br />

THE ACTION MAY CAUSE DAMAGE OR INJURY.<br />

IF THE USER PRESSES ABORT, THE SYSTEM<br />

RETURNS TO THE PREVIOUS SCROLL.<br />

A.9.13. Test Current Outputs<br />

WARNING<br />

EQUIPMENT MAY START<br />

CONTINUE ABORT<br />

- TEST SCROLL 5A -<br />

Current output #1<br />

should be 00.0 mA<br />

ENTER CLEAR<br />

Default: 4.0 mA<br />

Min: 0.0 mA<br />

Max: 20.0 mA<br />

To force the output, enter the desired number of milliamps and press ENTER.<br />

Press CLEAR to free the mA channel.<br />

If the board has additional optional boards, the following screen is shown.<br />

- TEST SCROLL 6 -<br />

Current output #2<br />

should be 00.0 mA<br />

ENTER CLEAR<br />

Default: 4.0 mA<br />

Min: 0.0 mA<br />

Max: 20.0 mA<br />

A.9.14. Test Current Inputs<br />

The following screen is displayed only if an analog input board is detected, and<br />

shows the status of each analog input channel.<br />

- TEST SCROLL 7 -<br />

Current input<br />

#1 00.0 V<br />

#2 00.0 V<br />

Password: Service<br />

<strong>Thermo</strong> Fisher Scientific Weighbridge Physical Parameters A-47<br />

REC 4175 Rev B